Linear Accuracy of Cone Beam CT Derived 3D Images

Overview
Journal Angle Orthod
Specialty Dentistry
Date 2009 Jan 7
PMID 19123719
Citations 57
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To compare the in vitro reliability and accuracy of linear measurements between cephalometric landmarks on cone beam computed tomography (CBCT) 3D volumetric images with varying basis projection images to direct measurements on human skulls.

Materials And Methods: Sixteen linear dimensions between 24 anatomic sites marked on 19 human skulls were directly measured. The skulls were imaged with CBCT (i-CAT, Imaging Sciences International, Hatfield, Pa) at three settings: (a) 153 projections, (b) 306 projections, and (c) 612 projections. The mean absolute error and modality mean (+/- SD) of linear measurements between landmarks on volumetric renderings were compared to the anatomic truth using repeated measures general linear model (P < or = .05).

Results: No difference in mean absolute error between the scan settings was found for almost all measurements. The average skull absolute error between marked reference points was less than the distances between unmarked reference sites. CBCT resulted in lower measurements for nine dimensions (mean difference range: 3.1 mm +/- 0.12 mm to 0.56 mm +/- 0.07 mm) and a greater measurement for one dimension (mean difference 3.3 mm +/- 0.12 mm). No differences were detected between CBCT scan sequences.

Conclusions: CBCT measurements were consistent between scan sequences and for direct measurements between marked reference points. Reducing the number of projections for 3D reconstruction did not lead to reduced dimensional accuracy and potentially provides reduced patient radiation exposure. Because the fiducial landmarks on the skulls were not radio-opaque, the inaccuracies found in measurement could be due to the methods applied rather than to innate inaccuracies in the CBCT scan reconstructions or 3D software employed.

Citing Articles

Definition of a coordinate system for multi-modal images of the temporal bone and inner ear.

Copson B, Wijewickrema S, Slinger C, Youssef D, Gerard J, OLeary S PLoS One. 2024; 19(10):e0294828.

PMID: 39374254 PMC: 11458053. DOI: 10.1371/journal.pone.0294828.


Three-dimensional assessment of low-level laser therapy on orthodontic miniscrew displacement using CBCT: a retrospective study.

Elfouly D, Abdel-Haffiez S, El-Harouni N, Elzoheiry M, Marzouk E BMC Oral Health. 2024; 24(1):998.

PMID: 39182096 PMC: 11344972. DOI: 10.1186/s12903-024-04711-x.


Correlation between cone-beam computed tomographic findings and the apnea-hypopnea index in obstructive sleep apnea patients: A cross-sectional study.

Isaac M, ElBeshlawy D, ElSobki A, Ahmed D, Kenawy S Imaging Sci Dent. 2024; 54(2):147-157.

PMID: 38948186 PMC: 11211029. DOI: 10.5624/isd.20230249.


Comparison of subjective image analysis and effective dose between low-dose cone-beam computed tomography machines.

Charuakkra A, Mahasantipiya P, Lehtinen A, Koivisto J, Jarnstedt J Dentomaxillofac Radiol. 2022; 52(2):20220176.

PMID: 36168973 PMC: 9974239. DOI: 10.1259/dmfr.20220176.


Cephalometric measurements performed on CBCT and reconstructed lateral cephalograms: a cross-sectional study providing a quantitative approach of differences and bias.

Baldini B, Cavagnetto D, Baselli G, Sforza C, Tartaglia G BMC Oral Health. 2022; 22(1):98.

PMID: 35351080 PMC: 8966183. DOI: 10.1186/s12903-022-02131-3.