» Articles » PMID: 19104166

Manual In-line Stabilization Increases Pressures Applied by the Laryngoscope Blade During Direct Laryngoscopy and Orotracheal Intubation

Overview
Journal Anesthesiology
Specialty Anesthesiology
Date 2008 Dec 24
PMID 19104166
Citations 25
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Manual in-line stabilization (MILS) is recommended during direct laryngoscopy and intubation in patients with known or suspected cervical spine instability. Because MILS impairs glottic visualization, the authors hypothesized that anesthesiologists would apply greater pressure during intubations with MILS than without.

Methods: Nine anesthetized and pharmacologically paralyzed patients underwent two sequential laryngoscopies and intubations, one with MILS and one without, in random order. A transducer array along a Macintosh 3 laryngoscope blade continuously measured applied pressures, and glottic view was characterized.

Results: With MILS, glottic visualization was worse in six patients, and intubation failure occurred in two of these six patients. Maximum laryngoscope pressure at best glottic view was greater with MILS than without (717 +/- 339 mmHg vs. 363 +/- 121 mmHg, respectively; n = 8; P = 0.023). Other measures of pressure application also indicated comparable increases with MILS.

Conclusion: Pressures applied to airway tissues by the laryngoscope blade are secondarily transmitted to the cervical spine and result in cranio-cervical motion. In the presence of cervical instability, impaired glottic visualization and secondary increases in pressure application with MILS have the potential to increase pathologic cranio-cervical motion.

Citing Articles

Standard Gum Elastic Bougie Versus Flexible Tip Bougie: Evaluation of Airway Adjuncts for Intubation by Paramedics in Entrapped Manikins with Difficult Airway Access-A Randomised, Controlled Trial.

Aleksandrowicz D, Ratajczyk P Healthcare (Basel). 2024; 12(22).

PMID: 39595430 PMC: 11593639. DOI: 10.3390/healthcare12222232.


Comparison of Efficacy of Intubation with HugeMed Non-channelled Video Laryngoscope, McCoy and Macintosh Laryngoscope in Simulated Difficult Laryngoscopy Using Rigid Cervical Collar: A Prospective Randomized Trial.

Murdeshwar G, Yashodha V, Poolandevi K, Malini S Ann Afr Med. 2024; 23(3):358-364.

PMID: 39034559 PMC: 11364341. DOI: 10.4103/aam.aam_193_23.


Airway management in patients with suspected or confirmed traumatic spinal cord injury: a narrative review of current evidence.

Wiles M Anaesthesia. 2022; 77(10):1120-1128.

PMID: 36089854 PMC: 9546380. DOI: 10.1111/anae.15807.


Success Rate on Endotracheal Intubation with Prone versus Kneeling Position in Mannequin Model with Limitation of Neck Movement: A Cross Over Study.

Tangkulpanich P, Jenpanitpong C, Patchkrua J, Silarak C, Srinaowech N, Thiamdao N Open Access Emerg Med. 2022; 14:177-182.

PMID: 35469276 PMC: 9034881. DOI: 10.2147/OAEM.S360169.


Airway management in maxillofacial trauma.

Saini S, Singhal S, Prakash S J Anaesthesiol Clin Pharmacol. 2021; 37(3):319-327.

PMID: 34759538 PMC: 8562439. DOI: 10.4103/joacp.JOACP_315_19.