Patient Exposure and Image Quality of Low-dose Pulmonary Computed Tomography Angiography: Comparison of 100- and 80-kVp Protocols

Overview
Journal Invest Radiol
Specialty Radiology
Date 2008 Nov 13
PMID 19002059
Citations 40
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: Measures to reduce radiation exposure and injected iodine mass are becoming more important with the widespread and often repetitive use of pulmonary CT angiography (CTA) in patients with suspected pulmonary embolism. In this retrospective study, we analyzed the capability of 2 low-kilovoltage CTA-protocols to achieve these goals.

Materials And Methods: Ninety patients weighing less than 100 kg were examined by a pulmonary CTA protocol using either 100 kVp (group A) or 80 kVp (group B). Volume and flow rate of contrast medium were reduced in group B (75 mL at 3 mL/s) compared with group A (100 mL at 4 mL/s). Attenuation was measured in the central and peripheral pulmonary arteries, and the contrast-to-noise ratios (CNR) were calculated. Entrance skin dose was estimated by measuring the surface dose in an ovoid-cylindrical polymethyl methacrylate chest phantom with 2 various dimensions corresponding to the range of chest diameters in our patients. Quantitative image parameters, estimated effective dose, and skin dose in both groups were compared by the t test. Arterial enhancement, noise, and overall quality were independently assessed by 3 radiologists, and results were compared between the groups using nonparametric tests.

Results: Mean attenuation in the pulmonary arteries in group B (427.6 +/- 116 HU) was significantly higher than in group A (342.1 +/- 87.7 HU; P < 0.001), whereas CNR showed no difference (group A, 20.6 +/- 7.3 and group B, 22.2 +/- 7.1; P = 0.302). Effective dose was lower by more than 40% with 80 kVp (1.68 +/- 0.23 mSv) compared with 100 kVp (2.87 +/- 0.88 mSv) (P < 0.001). Surface dose was significantly lower at 80 kVp compared with 100 kVp at both phantom dimensions (2.75 vs. 3.22 mGy; P = 0.027 and 2.22 vs. 2.73 mGy; P = 0.005, respectively). Image quality did not differ significantly between the groups (P = 0.151).

Conclusions: Using 80 kVp in pulmonary CTA permits reduced patient exposure by 40% and CM volume by 25% compared with 100 kVp without deterioration of image quality in patients weighing less than 100 kg.

Citing Articles

Machine Learning-Based Objective Evaluation Model of CTPA Image Quality: A Multi-Center Study.

Sun Q, Liu Z, Ding T, Shi C, Hou N, Sun C Int J Gen Med. 2025; 18:997-1005.

PMID: 40026813 PMC: 11869754. DOI: 10.2147/IJGM.S510784.


A Study of CT-derived Radiation Dose Calculation in Lung Q-SPECT/CT Imaging.

Silov G, Boga F Mol Imaging Radionucl Ther. 2023; 32(3):214-220.

PMID: 37870282 PMC: 10600546. DOI: 10.4274/mirt.galenos.2023.55707.


Feasibility and accuracy of coronary artery calcium score on virtual non-contrast images derived from a dual-layer spectral detector CT: A retrospective multicenter study.

Yang P, Zhao R, Deng W, An S, Li Y, Sheng M Front Cardiovasc Med. 2023; 10:1114058.

PMID: 36937907 PMC: 10018184. DOI: 10.3389/fcvm.2023.1114058.


Deep Learning Versus Iterative Reconstruction for CT Pulmonary Angiography in the Emergency Setting: Improved Image Quality and Reduced Radiation Dose.

Lenfant M, Chevallier O, Comby P, Secco G, Haioun K, Ricolfi F Diagnostics (Basel). 2020; 10(8).

PMID: 32759874 PMC: 7460033. DOI: 10.3390/diagnostics10080558.


CTPA with a conventional CT at 100 kVp vs. a spectral-detector CT at 120 kVp: Comparison of radiation exposure, diagnostic performance and image quality.

Sauter A, Shapira N, Kopp F, Aichele J, Bodden J, Knipfer A Eur J Radiol Open. 2020; 7:100234.

PMID: 32420413 PMC: 7215101. DOI: 10.1016/j.ejro.2020.100234.