Cobalt Determination with FI-FAAS After On-line Sorbent Preconcentration Using 1-nitroso-2-naphthol
Overview
Authors
Affiliations
The suitability of 1-nitroso-2-naphthol (NN) as a complexing agent for on-line preconcentration of cobalt using C(18) microcolumn with FI-FAAS system has been tested. Various parameters affecting the complex formation and its elution were optimized. Reagent solution (2.5x10(-3) mol l(-1)) and aqueous sample solution acidified with 0.1% (v/v) nitric acid were on-line mixed in a reaction coil set at 65+/-1 degrees C and flowed through the microcolumn for 30 s. The pH of the mixed solution was adjusted to 3 approximately 4 by adding HNO(3) (1 mol l(-1)) or NaOH (1 mol l(-1)) in the reagent solution. The adsorbed complexes in the microcolumn were eluted with ethanol (acidified to 1% nitric acid) in 10 s into the nebulizer of FAAS. A good precision (1.6% for 100 mug l(-1) Co(II), n=10), high enrichment factor 17.2, with detection limit (3sigma) 3.2 mug l(-1), and sample throughput (90 h(-1)) were obtained. The method was applied on the certified reference materials (CRMs) i.e. NBS-362 and NBS-364 (special low alloy steel), for the determination of cobalt and the results were in good agreement with the certified values.
Gopalakrishnan A, Angamaly S, Pradeep S, Madhusoodhanan D, Manoharan D, Mohanan P J Fluoresc. 2021; 32(1):189-202.
PMID: 34687395 DOI: 10.1007/s10895-021-02839-5.
Wang L, Chen Y, Zhang Z, Chen Y, Deng Q, Wang S Mikrochim Acta. 2021; 188(5):167.
PMID: 33877439 DOI: 10.1007/s00604-021-04813-1.
Tabaraki R, Sadeghinejad N, Nateghi A J Fluoresc. 2017; 28(1):251-257.
PMID: 29124631 DOI: 10.1007/s10895-017-2187-y.