» Articles » PMID: 18924837

A Submersible Battery-powered Flow Injection (FI) Sensor for the Determination of Nitrate in Estuarine and Coastal Waters

Overview
Date 2008 Oct 18
PMID 18924837
Authors
Affiliations
Soon will be listed here.
Abstract

The design, construction and performance of a remotely deployed submersible flow injection-based nutrient (total oxidized nitrogen) sensor are described. The sensor featured a custom-built microcomputer and a solid-state, flow-through spectrophotometric detector, and the derivatization chemistry was based on in-line copper-cadmium reduction of nitrate to nitrite, and diazotization with N1NED and sulphanilamide. The limit of detection was 0.0014 mg l(-1) NO3-N and the linear range was 0.0014- 0.77 mg l(-1) with a 260 microl sample volume and a 20 mm path length flow cell. Results from submersed deployments in the Tamar estuary and North Sea are also reported.