» Articles » PMID: 18849503

Mitral Valve and Tricuspid Valve Blood Flow: Accurate Quantification with 3D Velocity-encoded MR Imaging with Retrospective Valve Tracking

Overview
Journal Radiology
Specialty Radiology
Date 2008 Oct 14
PMID 18849503
Citations 88
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To validate flow assessment performed with three-dimensional (3D) three-directional velocity-encoded (VE) magnetic resonance (MR) imaging with retrospective valve tracking and to compare this modality with conventional two-dimensional (2D) one-directional VE MR imaging in healthy subjects and patients with regurgitation.

Materials And Methods: Patients and volunteers gave informed consent, and local medical ethics committee approval was obtained. Patient data were selected retrospectively and randomly from a database of MR studies obtained between July 2006 and July 2007. The 3D three-directional VE MR images were first validated in vitro and compared with 2D one-directional VE MR images. Mitral valve (MV) and tricuspid valve (TV) flow were assessed in 10 volunteers without valve insufficiency and 20 patients with valve insufficiency, with aortic systolic stroke volume (ASSV) as the reference standard.

Results: Phantom validation showed less than 5% error for both techniques. In volunteers, 3D three-directional VE MR images showed no bias for MV or TV flow when compared with ASSV, whereas 2D one-directional VE MR images showed significant bias for MV flow (15% overestimation, P < .01). TV flow showed 25% overestimation; however, this was insignificant because of the high standard deviation. Correlation with ASSV was strong for 3D three-directional VE MR imaging (r = 0.96, P < .01 for MV flow; r = 0.88, P < .01 for TV flow) and between MV and TV flow (r = 0.91, P < .01); however, correlation was weaker for 2D one-directional VE MR imaging (r = 0.80, P < .01 for MV flow; r = 0.22, P = .55 for TV flow) and between MV flow and TV flow (r = 0.34, P = .34). In patients (mean regurgitation fractions of 13% and 10% for MV flow and TV flow, respectively), correlation between MV flow and TV flow for 3D three-directional VE MR imaging was strong (r = 0.97, P < .01).

Conclusion: Use of 3D three-directional VE MR imaging enables accurate MV and TV flow quantification, even in patients with valve regurgitation.

Citing Articles

A Novel Valveless Pulsatile Flow Pump for Extracorporeal Blood Circulation.

Anatol J, Vignali E, Gasparotti E, Castro-Ruiz F, Rubio M, Barrios-Collado C Ann Biomed Eng. 2025; .

PMID: 39865185 DOI: 10.1007/s10439-024-03672-0.


Multimodality Imaging Evaluation of Diseases of the Pulmonic Valve and Right Ventricular Outflow Tract for the Adult Cardiologist.

Murphy S, Sultana S, Zern E, Tower-Rader A, Churchill J, Stefanescu Schmidt A Circ Cardiovasc Imaging. 2025; 18(2):e017126.

PMID: 39801475 PMC: 11835535. DOI: 10.1161/CIRCIMAGING.124.017126.


Assessment of Tricuspid Regurgitation by Cardiac Magnetic Resonance Imaging: Current Role and Future Applications.

Pype L, Domenech-Ximenos B, Paelinck B, Sturkenboom N, Van De Heyning C J Clin Med. 2024; 13(15).

PMID: 39124748 PMC: 11312898. DOI: 10.3390/jcm13154481.


Importance of Cardiovascular Magnetic Resonance Applied to Congenital Heart Diseases in Pediatric Age: A Narrative Review.

Moscatelli S, Pozza A, Leo I, Ielapi J, Scatteia A, Piana S Children (Basel). 2024; 11(7).

PMID: 39062326 PMC: 11276187. DOI: 10.3390/children11070878.


Tricuspid valve flow measurement using a deep learning framework for automated valve-tracking 2D phase contrast.

Lamy J, Gonzales R, Xiang J, Seemann F, Huber S, Steele J Magn Reson Med. 2024; 92(5):1838-1850.

PMID: 38817154 PMC: 11341256. DOI: 10.1002/mrm.30163.