Abstract

A magnetic particles (MPs)-based chemiluminescence immunoassay (CLIA) with high sensitivity, specificity, and reproducibility was proposed for the evaluation of estradiol (E(2)) in human sera. The MPs coated with secondary antibody were used as dispersed solid phase for the immunoassay, and the horseradish peroxidase (HRP)-luminol-H(2)O(2) chemiluminescent system with high sensitivity was chosen as the detection system. The method showed specific recognition to E(2), without cross-reaction for the major steroids, including estrone (E(1)), estriol (E(3)), dihydrotestosterone (DHT), androstenedione, and testosterone (T), which was commonly found in human serum. The addition of sodium trichloracetate (Na-TCA) in the enzyme buffer as a blocking agent contributed to the realization of direct analysis of E(2) in human serum without extraction. Besides, the effects of several physicochemical parameters, including the dilution ratios of E(2)-6-HRP conjugate and anti-E(2) polyclonal antibody, immunoreaction time, chemiluminescent (CL) substrate volume, volume of MPs, and CL reaction time, were studied and optimized. The proposed method had a detection limit of 2.51pgmL(-1) with a larger working range of 15-1000pgmL(-1). The inter-assay and intra-assay coefficient of variation (CV) were both less than 15%. The average recoveries of three different spiked concentration samples were 93.3, 106 and 101%, respectively. The method has been successfully applied to the determination of E(2) in 105 human sera and showed a good correlation compared with the commercial radioimmunoassay (RIA) kit with a correlative coefficient of 0.9892. This method has exhibited great potential in the fabrication of diagnostic kit and could be used in the clinical analysis of E(2) in human serum.

Citing Articles

Design and Validation of a Short Novel Estradiol Aptamer and Exploration of Its Application in Sensor Technology.

Jin H, Cheng Y, Kong F, Huang H, Yang Z, Wang X Molecules. 2024; 29(2).

PMID: 38276613 PMC: 10819485. DOI: 10.3390/molecules29020535.


A chemiluminescent magnetic microparticle immunoassay for the detection of antibody against African swine fever virus.

Shi Z, Cao L, Luo J, Zhou G, Zuo Q, Liu X Appl Microbiol Biotechnol. 2023; 107(11):3779-3788.

PMID: 37099055 DOI: 10.1007/s00253-023-12518-z.


Association between urine and serum estradiol levels in in vitro fertilization cycles.

Chotboon C, Salang L, Buppasiri P, Amnatbuddee S, Eamudomkarn N Sci Rep. 2022; 12(1):4393.

PMID: 35292698 PMC: 8924157. DOI: 10.1038/s41598-022-08292-z.


Determination of parvovirus antibodies in canine serum using magnetic bead-based chemiluminescence immunoassay.

Chen C, Guo X, Liang H, Ning B, Li J, Zhong S Biotechnol Appl Biochem. 2019; 66(4):586-590.

PMID: 31039295 PMC: 7161764. DOI: 10.1002/bab.1758.


Enhancement of the Chemiluminescence Response of Enzymatic Reactions by Plasmonic Surfaces for Biosensing Applications.

Abel B, Odukoya B, Mohammed M, Aslan K Nano Biomed Eng. 2015; 7(3):92-101.

PMID: 26582101 PMC: 4646423. DOI: 10.5101/nbe.v7i3.p92-101..