Commissioning and Quality Assurance of RapidArc Radiotherapy Delivery System

Overview
Specialties Oncology
Radiology
Date 2008 Sep 17
PMID 18793960
Citations 81
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: The Varian RapidArc is a system for intensity-modulated radiotherapy (IMRT) treatment planning and delivery. RapidArc incorporates capabilities such as variable dose-rate, variable gantry speed, and accurate and fast dynamic multileaf collimators (DMLC), to optimize dose conformality, delivery efficiency, accuracy and reliability. We developed RapidArc system commissioning and quality assurance (QA) procedures.

Methods And Materials: Tests have been designed that evaluate RapidArc performance in a stepwise manner. First, the accuracy of DMLC position during gantry rotation is examined. Second, the ability to vary and control the dose-rate and gantry speed is evaluated. Third, the combined use of variable DMLC speed and dose-rate is studied.

Results: Adapting the picket fence test for RapidArc, we compared the patterns obtained with stationary gantry and in RapidArc mode, and showed that the effect of gantry rotation on leaf accuracy was minimal (< or =0.2 mm). We then combine different dose-rates (111-600 MU/min), gantry speeds (5.5-4.3 degrees /s), and gantry range (Deltatheta = 90-12.9 degrees ) to give the same dose to seven parts of a film. When normalized to a corresponding open field (to account for flatness and asymmetry), the dose of the seven portions show good agreement, with a mean deviation of 0.7%. In assessing DMLC speed (0.46, 0.92, 1.84, and 2.76 cm/s) during RapidArc, the analysis of designed radiation pattern indicates good agreement, with a mean deviation of 0.4%.

Conclusions: The results of these tests provide strong evidence that DMLC movement, variable dose-rates and gantry speeds can be precisely controlled during RapidArc.

Citing Articles

Radiobiological analysis of VMAT treatment plan with flattened and flattening filter free photon beam: an EUD and TCP based comparative study.

Saroj D, Yadav S, Paliwal N, Haldar S, Shende R, Gupta G Rep Pract Oncol Radiother. 2024; 29(1):77-89.

PMID: 39165604 PMC: 11333070. DOI: 10.5603/rpor.99100.


Development and Evaluation of Jaw Position Detection Method in Jaw-tracking Delivery Using an Electronic Portal Imaging Device Cine Mode.

Kawabata T, Sugimoto S, Kurokawa C, Usui K, Inoue T, Nagata H Juntendo Iji Zasshi. 2024; 68(6):590-598.

PMID: 39081382 PMC: 11284286. DOI: 10.14789/jmj.JMJ22-0014-OA.


Improvement of treatment plan quality with modified fixed field volumetric modulated arc therapy in cervical cancer.

Jindakan S, Tharavichitkul E, Watcharawipha A, Nobnop W J Appl Clin Med Phys. 2024; 25(10):e14479.

PMID: 39032169 PMC: 11466474. DOI: 10.1002/acm2.14479.


[Study on the sensitivity of a volumetric modulated arc therapy plan verification equipment on multi-leaf collimator opening and closing errors and its gamma pass rate limit].

Hu J, Zou L, Gu S, Wang N, Cui F, Zhang S Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2023; 40(1):133-140.

PMID: 36854558 PMC: 9989765. DOI: 10.7507/1001-5515.202112018.


Dose rate versus gantry speed performance evaluation for slow gantry speeds using DICOM RT plans.

Bredikin A, Walsh M J Appl Clin Med Phys. 2022; 23(10):e13786.

PMID: 36121120 PMC: 9588258. DOI: 10.1002/acm2.13786.