Factors Associated with Failed Retrieval of the Günther Tulip Inferior Vena Cava Filter
Overview
Radiology
Authors
Affiliations
Purpose: To identify potential factors associated with failed retrieval of the Günther Tulip inferior vena cava (IVC) filter.
Materials And Methods: A retrospective review was performed of patients who underwent placement of the Günther Tulip filter with at least one attempt at filter retrieval over a 3-year period. Patient demographics, filter dwell time, filter angulation, and filter leg protrusion were analyzed.
Results: A total of 188 patients were included in the study. Primary retrieval success was achieved in 166 patients (88.3%), for an overall retrieval success rate of 94.2%. The overall mean dwell time was 63 days, whereas the mean dwell time in cases of retrieval failure was 95.4 days. A total of seven filters were in place for longer than 6 months, four of which were successfully retrieved. The degree of filter tilt was not found to be significantly related to retrieval success (P = .36), even though filter angulation was commonly cited as a reason for retrieval failure. On venography, 90.9% of filters that could not be retrieved showed leg protrusion beyond the lumen of the IVC. Finally, increasing patient age also correlated with retrieval failure (P = .01).
Conclusions: Prolonged dwell time and increasing patient age are associated with failed filter retrieval. However, even filters in place for extended periods can be safely removed.
Inferior Vena Cava Filter Retrieval: Simple to Complex.
Swersky A, Desai K Cardiovasc Intervent Radiol. 2024; 47(12):1642-1649.
PMID: 38396083 DOI: 10.1007/s00270-024-03673-5.
Eid F, Najim M, Elbanna M, Mostafa M, Magdi M Eur J Case Rep Intern Med. 2024; 11(1):004192.
PMID: 38223284 PMC: 10783465. DOI: 10.12890/2023_004192.
The application of inferior vena cava filters in orthopaedics and current research advances.
He J, Wang Z, Zhou Y, Ni H, Sun X, Xue J Front Bioeng Biotechnol. 2022; 10:1045220.
PMID: 36479430 PMC: 9719953. DOI: 10.3389/fbioe.2022.1045220.
White P, Jahangiri Y, Farsad K, Kaufman J, Al-Hakim R Diagn Interv Radiol. 2021; 27(5):644-648.
PMID: 34559049 PMC: 8480951. DOI: 10.5152/dir.2021.19411.
Cardiac Tamponade Secondary to IVC Filter Retrieval.
Margolin E, Kwan W, Kanjanauthai S, Sohn J JACC Case Rep. 2021; 2(6):873-876.
PMID: 34317371 PMC: 8302023. DOI: 10.1016/j.jaccas.2020.04.014.