» Articles » PMID: 18725095

Factors Associated with Failed Retrieval of the Günther Tulip Inferior Vena Cava Filter

Overview
Date 2008 Aug 30
PMID 18725095
Citations 19
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To identify potential factors associated with failed retrieval of the Günther Tulip inferior vena cava (IVC) filter.

Materials And Methods: A retrospective review was performed of patients who underwent placement of the Günther Tulip filter with at least one attempt at filter retrieval over a 3-year period. Patient demographics, filter dwell time, filter angulation, and filter leg protrusion were analyzed.

Results: A total of 188 patients were included in the study. Primary retrieval success was achieved in 166 patients (88.3%), for an overall retrieval success rate of 94.2%. The overall mean dwell time was 63 days, whereas the mean dwell time in cases of retrieval failure was 95.4 days. A total of seven filters were in place for longer than 6 months, four of which were successfully retrieved. The degree of filter tilt was not found to be significantly related to retrieval success (P = .36), even though filter angulation was commonly cited as a reason for retrieval failure. On venography, 90.9% of filters that could not be retrieved showed leg protrusion beyond the lumen of the IVC. Finally, increasing patient age also correlated with retrieval failure (P = .01).

Conclusions: Prolonged dwell time and increasing patient age are associated with failed filter retrieval. However, even filters in place for extended periods can be safely removed.

Citing Articles

Inferior Vena Cava Filter Retrieval: Simple to Complex.

Swersky A, Desai K Cardiovasc Intervent Radiol. 2024; 47(12):1642-1649.

PMID: 38396083 DOI: 10.1007/s00270-024-03673-5.


Cardiac Arrest Following Retrieval of Inferior Vena Cava Filter: A Case Report and Literature Review of Pericardial Effusion and Cardiac Tamponade.

Eid F, Najim M, Elbanna M, Mostafa M, Magdi M Eur J Case Rep Intern Med. 2024; 11(1):004192.

PMID: 38223284 PMC: 10783465. DOI: 10.12890/2023_004192.


The application of inferior vena cava filters in orthopaedics and current research advances.

He J, Wang Z, Zhou Y, Ni H, Sun X, Xue J Front Bioeng Biotechnol. 2022; 10:1045220.

PMID: 36479430 PMC: 9719953. DOI: 10.3389/fbioe.2022.1045220.


Lateral tilt during IVC filter placement does not predict the need for advanced filter retrieval techniques.

White P, Jahangiri Y, Farsad K, Kaufman J, Al-Hakim R Diagn Interv Radiol. 2021; 27(5):644-648.

PMID: 34559049 PMC: 8480951. DOI: 10.5152/dir.2021.19411.


Cardiac Tamponade Secondary to IVC Filter Retrieval.

Margolin E, Kwan W, Kanjanauthai S, Sohn J JACC Case Rep. 2021; 2(6):873-876.

PMID: 34317371 PMC: 8302023. DOI: 10.1016/j.jaccas.2020.04.014.