Wear Studies on the Likely Performance of CFR-PEEK/CoCrMo for Use As Artificial Joint Bearing Materials

Overview
Publisher Springer
Date 2008 Aug 16
PMID 18704647
Citations 14
Authors
Affiliations
Soon will be listed here.
Abstract

It is well known that a reduction in the volume of wear produced by articulating surfaces in artificial joints is likely to result in a lower incidence of failure due to wear particle induced osteolysis. Therefore, new materials have been introduced in an effort to produce bearing surfaces with lower, more biologically acceptable wear. Polyetheretherketone (PEEK-OPTIMA) has been successfully used in a number of implant applications due to its combination of mechanical strength and biocompatibility. Pin-on-plate wear tests were performed on various combinations of PEEK-OPTIMA and carbon fibre reinforced PEEK-OPTIMA (CFR-PEEK) against various CoCrMo alloys to assess the potential of this material combination for use in orthopaedic implants. The PEEK/low carbon CoCrMo produced the highest wear. CFR-PEEK against high carbon or low carbon CoCrMo provided low wear factors. Pin-on-plate tests performed on ultra-high molecular weight polyethylene (UHMWPE) against CoCrMo (using comparable test conditions) have shown similar or higher wear than that found for CFR-PEEK/CoCrMo. This study gives confidence in the likelihood of this material combination performing well in orthopaedic applications.

Citing Articles

Effect of carbon-fiber-reinforced polyetheretherketone on stress distribution in a redesigned tumor-type knee prosthesis: a finite element analysis.

Wu H, Guo Y, Guo W Front Bioeng Biotechnol. 2023; 11:1243936.

PMID: 37823023 PMC: 10562634. DOI: 10.3389/fbioe.2023.1243936.


Consequences of using poly-ether-ether-ketone versus traditional implant on tibial cement penetration and short-term clinical outcomes during total knee arthroplasty: a randomized controlled trial.

Zhao G, Yao S, Sun X, Ma J, Wang J J Orthop Surg Res. 2023; 18(1):589.

PMID: 37559133 PMC: 10410788. DOI: 10.1186/s13018-023-04064-1.


Long Term Results of a Rotating Hinge Total Knee Prosthesis With Carbon-Fiber Reinforced Poly-Ether-Ether-Ketone (CFR-PEEK) as Bearing Material.

Vertesich K, Staats K, Bohler C, Koza R, Lass R, Giurea A Front Bioeng Biotechnol. 2022; 10:845859.

PMID: 35356773 PMC: 8959897. DOI: 10.3389/fbioe.2022.845859.


ZrN-ZrON vs ZrO-ZrON coatings deposited via unbalanced DC magnetron sputtering.

Cubillos G, Romero E, Umana-Perez A Sci Rep. 2021; 11(1):18926.

PMID: 34556694 PMC: 8460672. DOI: 10.1038/s41598-021-98052-2.


Polyetheretherketone and Its Composites for Bone Replacement and Regeneration.

Liao C, Li Y, Tjong S Polymers (Basel). 2020; 12(12).

PMID: 33260490 PMC: 7760052. DOI: 10.3390/polym12122858.


References
1.
Sagomonyants K, Jarman-Smith M, Devine J, Aronow M, Gronowicz G . The in vitro response of human osteoblasts to polyetheretherketone (PEEK) substrates compared to commercially pure titanium. Biomaterials. 2008; 29(11):1563-72. DOI: 10.1016/j.biomaterials.2007.12.001. View

2.
Gioe T, Novak C, Sinner P, Ma W, Mehle S . Knee arthroplasty in the young patient: survival in a community registry. Clin Orthop Relat Res. 2007; 464:83-7. DOI: 10.1097/BLO.0b013e31812f79a9. View

3.
Tipper J, Firkins P, Ingham E, Fisher J, Stone M, Farrar R . Quantitative analysis of the wear and wear debris from low and high carbon content cobalt chrome alloys used in metal on metal total hip replacements. J Mater Sci Mater Med. 2004; 10(6):353-62. DOI: 10.1023/a:1026473723777. View

4.
Roberts V, Esler C, Harper W . A 15-year follow-up study of 4606 primary total knee replacements. J Bone Joint Surg Br. 2007; 89(11):1452-6. DOI: 10.1302/0301-620X.89B11.19783. View

5.