Near-patient Testing of Potassium Levels Using Arterial Blood Gas Analysers: Can We Trust These Results?

Overview
Journal Emerg Med J
Specialty Emergency Medicine
Date 2008 Jul 29
PMID 18660404
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Near-patient testing allows rapid availability of results to enable prompt decision-making. Potassium abnormalities are common in acutely ill patients and can be associated with life-threatening complications. At times there is uncertainty whether clinical decisions can be based on the potassium result obtained from arterial blood gas (ABG) analysers or if laboratory values should be awaited.

Objectives: To determine the opinion of clinicians regarding the use of blood gas analysers to measure potassium and to determine the level of agreement between blood gas analyser and laboratory measurements of potassium in arterial blood samples.

Method: Survey of 64 doctors using a questionnaire and a retrospective comparative study of 529 paired results of ABG and arterial laboratory measurements of potassium in 121 critically ill patients.

Results: 51.6% of the doctors would wait for laboratory confirmation and 48.4% would base clinical decisions on results obtained from the blood gas analyser. The difference between the means of potassium values from the two methods is 0.03 mmol/l (95% CI 0.011 to 0.056; p = 0.0041). The 95% limits of agreement were from -0.485 mmol/l (95% CI -0.524 to -0.447) to 0.551 mmol/l (95% CI 0.513 to 0.590). 95% of the results fell within the difference limits of 0.5 mmol/l.

Conclusions: Most clinicians still await laboratory confirmation of results obtained from blood gas analysers but in this setting there is sufficient agreement between the results obtained from the authors' blood gas analyser and a laboratory analyser to enable effective clinical decisions to be made.

Citing Articles

Ionised hypocalcaemia in emergency and acute medicine.

Yates G, Barrett A, Ogedengbe O BMJ Case Rep. 2022; 15(11).

PMID: 36375853 PMC: 9664304. DOI: 10.1136/bcr-2022-251611.


Automatic real-time analysis and interpretation of arterial blood gas sample for Point-of-care testing: Clinical validation.

Rodriguez-Villar S, Poza-Hernandez P, Freigang S, Zubizarreta-Ormazabal I, Paz-Martin D, Holl E PLoS One. 2021; 16(3):e0248264.

PMID: 33690724 PMC: 7946183. DOI: 10.1371/journal.pone.0248264.


Are Serum Potassium and Magnesium Levels Associated with Atrial Fibrillation After Cardiac Surgery?.

Howitt S, Grant S, Campbell N, Malagon I, McCollum C J Cardiothorac Vasc Anesth. 2020; 34(5):1152-1159.

PMID: 31948890 PMC: 7610845. DOI: 10.1053/j.jvca.2019.10.045.


Determination of Electrolytes in Critical Illness Patients at Different pH Ranges: Whom Shall We Believe, the Blood Gas Analysis or the Laboratory Autoanalyzer?.

Hohmann C, Pfister R, Kuhr K, Merkle J, Hinzmann J, Michels G Crit Care Res Pract. 2019; 2019:9838706.

PMID: 31396417 PMC: 6664556. DOI: 10.1155/2019/9838706.


Interchangeability of Sodium and Potassium Result Values of Arterial Blood Gas with Laboratory Analyzer: Narrative Review.

Ustundag Y, Huysal K, Ozgunay S, Turkoglu A Indian J Crit Care Med. 2019; 23(1):35-42.

PMID: 31065207 PMC: 6481262. DOI: 10.5005/jp-journals-10071-23110.