Retrospective Analysis of Adverse Events in Preclinical Ventricular Assist Device Experiments

Overview
Journal ASAIO J
Specialty General Surgery
Date 2008 Jul 23
PMID 18645350
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Ventricular assist devices (VADs) are a widely acceptable therapeutic option for patients with end-stage heart failure. Data from preclinical animal trials provide important information about the efficacy, biocompatibility, and functioning of these devices. However, animal experiments have several limitations that may significantly affect outcomes even after implanting a properly functioning device. This retrospective analysis was performed to analyze, at explant, adverse events associated with VADs during the animal phase of testing. Data were collected from 141 VAD experiments performed in cattle that had received 11 different types of devices. A total of 294 adverse events were documented. Inflow cannula problems were cited in 79 implants (56%) and pump-related issues in 55 cases (39%). Adverse events associated with connectors were found in 53 implants (38%). Adverse events directly having to do with the animals represented only 17 implants (12%) in this series. These findings suggest that preclinical testing needs to place greater emphasis on the entire VAD system. In support of these results, human clinical data seem to indicate that actual VAD failure is rarely a cause for premature device removal.

Citing Articles

Effect of flow rate ratio and positioning on a lighthouse tip ECMO return cannula.

Fiusco F, Lemetayer J, Broman L, Prahl Wittberg L Biomech Model Mechanobiol. 2023; 22(6):1891-1899.

PMID: 37454305 PMC: 10613146. DOI: 10.1007/s10237-023-01741-2.


Numerical and experimental investigation of a lighthouse tip drainage cannula used in extracorporeal membrane oxygenation.

Fiusco F, Rorro F, Broman L, Prahl Wittberg L Artif Organs. 2022; 47(2):330-341.

PMID: 36227654 PMC: 10092507. DOI: 10.1111/aor.14421.


In Vitro and In Vivo Performance Evaluation of the Second Developmental Version of the PediaFlow Pediatric Ventricular Assist Device.

Maul T, Kocyildirim E, Johnson Jr C, Daly A, Olia S, Woolley J Cardiovasc Eng Technol. 2012; 2(4):253-262.

PMID: 22211150 PMC: 3247142. DOI: 10.1007/s13239-011-0061-7.


Platelet activation in ovines undergoing sham surgery or implant of the second generation PediaFlow pediatric ventricular assist device.

Johnson Jr C, Wearden P, Kocyildirim E, Maul T, Woolley J, Ye S Artif Organs. 2011; 35(6):602-13.

PMID: 21463346 PMC: 3178345. DOI: 10.1111/j.1525-1594.2010.01124.x.


Computational fluid dynamics analysis of thrombosis potential in left ventricular assist device drainage cannulae.

Fraser K, Zhang T, Taskin M, Griffith B, Wu Z ASAIO J. 2010; 56(3):157-63.

PMID: 20400890 PMC: 2875940. DOI: 10.1097/MAT.0b013e3181d861f1.