Reference Dosimetry in a Scanned Pulsed Proton Beam Using Ionisation Chambers and a Faraday Cup

Overview
Journal Phys Med Biol
Publisher IOP Publishing
Date 2008 Jun 17
PMID 18552418
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

In order to give the correct dose to a patient, the monitor chamber for a proton scanning system has to be calibrated. As recombination of ion pairs occurs in the monitor chamber, the relation between the number of particles traversing it per time unit and the ionization chamber signal is not linear. A method developed for a scanned pulsed proton beam taking the nonlinear monitor signal into account is described. A vital part of the reference dosimetry procedure is to determine the absorbed dose under reference conditions, which is recommended to be done with an ionization chamber. For a scanned pulsed proton beam, the recombination in the ionization chamber is not negligible and the signal from the ionization chamber has to be corrected. In this work, it is shown that although the pulse length is comparable to the ion transit time the beam can be considered as continuously scanned if the applied high voltage is not too small. Also shown is that the two-voltage formula for a continuous beam is under some conditions applicable for a continuous scanned beam as well.

Citing Articles

ESTRO-EPTN radiation dosimetry guidelines for the acquisition of proton pencil beam modelling data.

Goma C, Henkner K, Jakel O, Lorentini S, Magro G, Mirandola A Phys Imaging Radiat Oncol. 2024; 31:100621.

PMID: 39220113 PMC: 11364130. DOI: 10.1016/j.phro.2024.100621.


Ultrahigh dose rate pencil beam scanning proton dosimetry using ion chambers and a calorimeter in support of first in-human FLASH clinical trial.

Lee E, Lourenco A, Speth J, Lee N, Subiel A, Romano F Med Phys. 2022; 49(9):6171-6182.

PMID: 35780318 PMC: 9546035. DOI: 10.1002/mp.15844.


Use of a novel two-dimensional ionization chamber array for pencil beam scanning proton therapy beam quality assurance.

Lin L, Kang M, Solberg T, Mertens T, Baeumer C, Ainsley C J Appl Clin Med Phys. 2015; 16(3):5323.

PMID: 26103492 PMC: 5690130. DOI: 10.1120/jacmp.v16i3.5323.


Commissioning dose computation models for spot scanning proton beams in water for a commercially available treatment planning system.

Zhu X, Poenisch F, Lii M, Sawakuchi G, Titt U, Bues M Med Phys. 2013; 40(4):041723.

PMID: 23556893 PMC: 3631269. DOI: 10.1118/1.4798229.


Commissioning of the discrete spot scanning proton beam delivery system at the University of Texas M.D. Anderson Cancer Center, Proton Therapy Center, Houston.

Gillin M, Sahoo N, Bues M, Ciangaru G, Sawakuchi G, Poenisch F Med Phys. 2010; 37(1):154-63.

PMID: 20175477 PMC: 11078095. DOI: 10.1118/1.3259742.