Insertional Torque and Axial Pull-out Strength of Mini-implants in Mandibles of Dogs

Overview
Publisher Elsevier
Specialty Dentistry
Date 2008 Jun 10
PMID 18538236
Citations 20
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: The objectives of this study were to evaluate insertional torque of mini-implants (MI) for orthodontic anchorage, to compare their axial pull-out strength at 0 (T0), 15 (T15), and 60 (T60) days after placement, to determine initial and peri-implant cortical bone thickness, and to analyze the correlations among these variables.

Methods: Sixty MI were placed in the mandibles of 10 dogs. Peak insertional torque values were recorded, and, after the animals were killed, specimens were prepared. The MI were aligned with a testing machine to register axial pull-out strength. Cortical bone thickness was measured after removal of the MI. Statistical analyses were conducted with ANOVA and Tukey HSD tests.

Results: Pull-out strength and peri-implant cortical bone thickness had differences (P <0.01) at the various times. The T0 value showed the highest pull-out strength, and the smallest peri-implant cortical bone thickness was at T60. There was a regular correlation between pull-out strength and peri-implant cortical bone thickness at T0 (r = 0.44, P = 0.05). There was no correlation between insertional torque and pull-out strength, and insertional torque and initial cortical bone thickness.

Conclusions: Pull-out strength is greater immediately after placement of MI, cortical bone thickness decreases because of bone resorption, and insertional torque is not an efficient method for predicting the retention of MI.

Citing Articles

Effects of exposure length, cortical and trabecular bone contact areas on primary stability of infrazygomatic crest mini-screws at different insertion angles.

Du B, Lin Y, Ji M, Yang Q, Jiang J, Wang F BMC Oral Health. 2024; 24(1):924.

PMID: 39123162 PMC: 11316306. DOI: 10.1186/s12903-024-04626-7.


Calcium phosphate cement promotes the stability of osteoporotic lumbar pedicle screw by enhancer-injecters with different number of holes.

Fu S, Zhang Y, Wang R, Zou X, Ai F, Wang J BMC Surg. 2023; 23(1):354.

PMID: 37980464 PMC: 10657603. DOI: 10.1186/s12893-023-02235-9.


Thread shape, cortical bone thickness, and magnitude and distribution of stress caused by the loading of orthodontic miniscrews: finite element analysis.

Sakamaki T, Watanabe K, Iwasa A, Deguchi T, Horiuchi S, Tanaka E Sci Rep. 2022; 12(1):12367.

PMID: 35859046 PMC: 9300621. DOI: 10.1038/s41598-022-16662-w.


Effects of the Healing Chambers in Implant Macrogeometry Design in a Low-Density Bone Using Conventional and Undersized Drilling.

Gehrke S, Scarano A, de Lima J, Bianchini M, Dedavid B, De Aza P J Int Soc Prev Community Dent. 2021; 11(4):437-447.

PMID: 34430506 PMC: 8352059. DOI: 10.4103/jispcd.JISPCD_96_21.


Development of a quantitative preclinical screening model for implant osseointegration in rat tail vertebra.

Farkasdi S, Pammer D, Racz R, Hriczo-Koperdak G, Szabo B, Dobo-Nagy C Clin Oral Investig. 2018; 23(7):2959-2973.

PMID: 30374828 PMC: 7398390. DOI: 10.1007/s00784-018-2661-1.