» Articles » PMID: 18496416

Theory and in Vitro Validation of a New Extracorporeal Arteriovenous Loop Approach for Hemodynamic Assessment in Pediatric and Neonatal Intensive Care Unit Patients

Overview
Specialty Pediatrics
Date 2008 May 23
PMID 18496416
Citations 19
Authors
Affiliations
Soon will be listed here.
Abstract

Objectives: No simple method exists for repeatedly measuring cardiac output in intensive care pediatric and neonatal patients. The purpose of this study is to present the theory and examine the in vitro accuracy of a new ultrasound dilution cardiac output measurement technology in which an extracorporeal arteriovenous tubing loop is inserted between existing arterial and venous catheters.

Design: Laboratory experiments.

Setting: Research laboratory.

Subjects: None.

Interventions: None.

Measurements And Main Results: In vitro validations of cardiac output, central blood volume, total end-diastolic volume, and active circulation volume were performed in a model mimicking pediatric (children 2-10 kg) and neonatal (0.5-3 kg) flows and volumes against flows and volumes measured volumetrically. Reusable sensors were clamped onto the arterial and venous limbs of the arteriovenous loop. A peristaltic pump was used to circulate liquid at 6-12 mL/min from the artery to the vein through the arteriovenous loop. Body temperature injections of isotonic saline (0.3-10 mL) were performed. In the pediatric setting, the absolute difference between cardiac output measured by dilution and cardiac output measured volumetrically was 3.97% +/- 2.97% (range 212-1200 mL/min); for central blood volume the difference was 4.59% +/- 3.14% (range 59-315 mL); for total end-diastolic volume the difference was 4.10% +/- 3.08% (range 24-211 mL); and for active circulation volume the difference was 3.30% +/- 3.07% (range 247-645 mL). In the neonatal setting the difference for cardiac output was 4.40% +/- 4.09% (range 106-370 mL/min); for central blood volume the difference was 4.90% +/- 3.69% (range 50-62 mL); and for active circulation volume the difference was 5.39% +/- 4.42% (range 104-247 mL).

Conclusions: In vitro validation confirmed the ability of the ultrasound dilution technology to accurately measure small flows and volumes required for hemodynamic assessments in small pediatric and neonatal patients. Clinical studies are in progress to assess the reliability of this technology under different clinical situations.

Citing Articles

Ultrasound dilution cardiac output and echocardiography findings in anesthetized mature alpacas (Vicugna pacos) during normotension, hypotension and hypertension.

Diaz-Falcon N, Clark-Price S, Holland M, Johnson J, Lascola K PLoS One. 2023; 18(4):e0284299.

PMID: 37036882 PMC: 10085030. DOI: 10.1371/journal.pone.0284299.


Indexing haemodynamic variables in young children.

Sigurdsson T, Lindberg L Acta Anaesthesiol Scand. 2020; 65(2):195-202.

PMID: 33015826 PMC: 7820964. DOI: 10.1111/aas.13720.


A modified method of measuring plasma volume with indocyanine green: reducing the frequency of blood sampling while maintaining accuracy.

Kim K, Park D, Kang E, Kim M, Choi B, Noh G J Clin Monit Comput. 2020; 35(4):779-785.

PMID: 32472337 DOI: 10.1007/s10877-020-00536-5.


Indirect Calorimetry Overestimates Oxygen Consumption in Young Children: Caution is Advised Using Direct Fick Method as a Reference Method in Cardiac Output Comparison Studies.

Sigurdsson T, Lindberg L Pediatr Cardiol. 2019; 41(1):149-154.

PMID: 31741015 PMC: 6987070. DOI: 10.1007/s00246-019-02238-5.


Cardiac output measurements via echocardiography versus thermodilution: A systematic review and meta-analysis.

Zhang Y, Wang Y, Shi J, Hua Z, Xu J PLoS One. 2019; 14(10):e0222105.

PMID: 31581196 PMC: 6776392. DOI: 10.1371/journal.pone.0222105.


References
1.
Wood E . Diagnostic applications of indicator-dilution technics in congenital heart disease. Circ Res. 1962; 10:531-68. DOI: 10.1161/01.res.10.3.531. View

2.
Krivitski N, Starostin D, Smith T . Extracorporeal recording of mouse hemodynamic parameters by ultrasound velocity dilution. ASAIO J. 1999; 45(1):32-6. DOI: 10.1097/00002480-199901000-00008. View

3.
Belorusov O, Kislukhin V, Krivitskii N, Maksimenko V, Shifris S . [Study of parameters of central hemodynamics in patients during hemodialysis]. Khirurgiia (Mosk). 1989; (6):21-3. View

4.
Kisch H, Leucht S, Lichtwarck-Aschoff M, Pfeiffer U . Accuracy and reproducibility of the measurement of actively circulating blood volume with an integrated fiberoptic monitoring system. Crit Care Med. 1995; 23(5):885-93. DOI: 10.1097/00003246-199505000-00017. View

5.
Krivitski N, Depner T . Cardiac output and central blood volume during hemodialysis: methodology. Adv Ren Replace Ther. 1999; 6(3):225-32. DOI: 10.1016/s1073-4449(99)70018-x. View