A Guide to the Removal of Heavy Silicone Oil

Overview
Journal Br J Ophthalmol
Specialty Ophthalmology
Date 2008 May 8
PMID 18460540
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Aim: Heavy silicone oil removal can be challenging and differs considerably from conventional oil. Traditionally, strong active aspiration had to be applied through a long 18G needle just above the optic disc. We present a novel technique using a much shorter (7.5 mm) and smaller (20G) needle allowing its removal "from a distance."

Method: Active aspiration on a vacuum of 600 mm Hg of the "viscous fluid injector" was applied using the 20G cannula in a polymethylmethacrylate model eye chamber that was surface-modified to mimic the surface properties of the retina. Measurements were taken using still photographs.

Results: Under injection the maximum diameter of a silicone oil bubble supported by interfacial tension alone was 5 mm for a steel and 7 mm for a polyurethane cannula. Under suction, the silicone bubble changed shape and became conical, thus further increasing the cannula's reach. This conical shape illustrated "tubeless siphoning," which is a physical property of non-Newtonian fluids.

Discussion: The use of shorter and smaller gauge cannula for removal of Densiron obviates the need to enlarge the sclerotomy beyond 20G or to apply suction in close proximity to disc and fovea. This potentially reduces the risk of iatrogenic damage such as entry site tears or postoperative hypotony.

Citing Articles

Sustained-release drug delivery systems.

Williams R, Cauldbeck H, Kearns V Eye (Lond). 2024; 39(4):658-666.

PMID: 38760462 PMC: 11885436. DOI: 10.1038/s41433-024-03134-w.


Surgical technique for removal of high-density silicone oil (Oxane HD).

Oliveira R, de Moraes N, Fernandes R, Magalhaes Jr O, Maia M Int J Retina Vitreous. 2023; 9(1):64.

PMID: 37872580 PMC: 10591402. DOI: 10.1186/s40942-023-00501-9.


Optimizing Retinal Thermofusion in Retinal Detachment Repair: Achieving Instant Adhesion without Air Tamponade.

Heriot W, Metha A, He Z, Lim J, Hoang A, Nishimura T Ophthalmol Sci. 2022; 2(4):100179.

PMID: 36531586 PMC: 9754973. DOI: 10.1016/j.xops.2022.100179.


Thermofusion of the retina with the RPE to seal tears during retinal detachment repair.

Heriot W Graefes Arch Clin Exp Ophthalmol. 2016; 254(4):691-6.

PMID: 26916781 DOI: 10.1007/s00417-016-3295-0.


F6H8 as an Intraoperative Tool and F6H8/Silicone Oil as a Postoperative Tamponade in Inferior Retinal Detachment with Inferior PVR.

Tosi G, Marigliani D, Bacci T, Romeo N, Balestrazzi A, Martone G J Ophthalmol. 2014; 2014:956831.

PMID: 24672710 PMC: 3941958. DOI: 10.1155/2014/956831.