» Articles » PMID: 18395820

Electrospinning Fabrication of High-trackable Catheter Tip with Gradually Graded or Gradient Flexibility

Overview
Date 2008 Apr 9
PMID 18395820
Citations 1
Authors
Affiliations
Soon will be listed here.
Abstract

Increasing demand for a flexible catheter tip exists in the rapidly growing fields of vascular and interventional radiology. This article describes a fabrication technology producing a catheter tip with gradually graded or gradient flexibility. Based on deflection mechanics of a tubular construct, three models were incorporated into structural designs. The models included graded wall thickness, tapered shape, or graded Young's moduli. Electrospinning using elastomeric polymer [poly(L-lactide-co-epsilon-caprolactone), polyurethane] solutions on a transversely and rotationally moving mandrel enabled preparation of three prototype tubular tips by either regio-specific control of wall thickness, tapered mandrel, or regiospecific deposition of polyurethanes with different Young's moduli. All prototype tips exhibited high flexibility at tip ends, as verified using a bifurcated vessel model.

Citing Articles

Three-Dimensional Melted Electrowriting Drug Coating Fibers for the Prevention of Device-Associated Infections: A Pilot Study.

Martins J, da Silva E, Fernandes A, Costa de Oliveira S Bioengineering (Basel). 2024; 11(7).

PMID: 39061718 PMC: 11273671. DOI: 10.3390/bioengineering11070636.