Comparison of the Heat Generation of Light Curing Units

Overview
Specialty Dentistry
Date 2008 Feb 12
PMID 18264527
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Aim: The aim of this study was to evaluate the heat generation of three different types of light curing units.

Methods And Materials: Temperature increases were recorded from a distance of 1 mm from a thermocouple to the tip of three different types of light curing units including one quartz-tungsten halogen (QTH), one plasma arc (PAC), and one light emitting diode (LED) unit. An experimental model was designed to fix the 1 mm distance between the tip of the light curing units and the thermocouple wire. Temperature changes were recorded in 10 second intervals up to 40 seconds. (10, 20, 30, and 40 seconds). Temperature measurements were repeated three times for every light curing unit after a one hour standby period. Statistical analysis of the results was performed using the analysis of variance (ANOVA) and the Bonferroni Test.

Results: The highest temperature rises (54.4+/-1.65 degrees C) occurred during activation of a PAC light curing unit for every test period (p<.05). The least temperature increase (11.8+/-1.3 degrees C) occurred with a LED curing unit for each tested period except for the measurement of the temperature rise using the QTH curing unit at the tenth second interval (p<.05).

Conclusion: These results indicate the choice of light activation unit and curing time is important when polymerizing light activated resin based restorations to avoid any thermal damage to the pulp.

Citing Articles

Temperatures in the pulpal cavity during orthodontic bonding using an LED light curing unit : An in vitro pilot study.

Groddeck A, Schwestka-Polly R, Hecker H, Sostmann M J Orofac Orthop. 2020; 81(6):396-406.

PMID: 32875349 DOI: 10.1007/s00056-020-00246-2.


Effects of High-Energy Curing Lights on Time-Dependent Temperature Changes of Pulp Space During Orthodontic Bonding.

Gul Amuk N, Kurt G, Er O, Cakmak G, Aslantas V Turk J Orthod. 2019; 32(1):22-27.

PMID: 30944896 PMC: 6436908. DOI: 10.5152/TurkJOrthod.2019.19015.


Influence of extended light exposure curing times on the degree of conversion of resin-based pit and fissure sealant materials.

Fatima N Saudi Dent J. 2014; 26(4):151-5.

PMID: 25382947 PMC: 4223770. DOI: 10.1016/j.sdentj.2014.05.002.


Influence of softening test and light-activation protocols on resin composite polymer structure.

Giorgi M, Lima D, Marchi G, Ambrosano G, Aguiar F Eur J Dent. 2014; 8(1):9-14.

PMID: 24966740 PMC: 4054039. DOI: 10.4103/1305-7456.126233.


Characterization of heat emission of light-curing units.

Wahbi M, Aalam F, Fatiny F, Radwan S, Eshan I, Al-Samadani K Saudi Dent J. 2013; 24(2):91-8.

PMID: 23960535 PMC: 3723364. DOI: 10.1016/j.sdentj.2012.01.003.