» Articles » PMID: 18215679

Effect of Electropolishing ProFile Nickel-titanium Rotary Instruments on Cyclic Fatigue Resistance, Torsional Resistance, and Cutting Efficiency

Overview
Journal J Endod
Publisher Elsevier
Specialty Dentistry
Date 2008 Jan 25
PMID 18215679
Citations 18
Authors
Affiliations
Soon will be listed here.
Abstract

This study investigated the effect of electropolishing ProFile nickel titanium rotary instruments on torque resistance, fatigue resistance, and cutting efficiency. Size 25/.04 ProFile files (Dentsply Maillefer, Switzerland) that were nonpolished for the control group (n = 15) and electropolished for the experimental group (n = 15) were used for each experiment. Cyclical fatigue was determined by counting rotations until breakage with an applied 30 degrees , 45 degrees , and 60 degrees curve with a 5-mm radius. Torque and angle at failure were measured by rotating clamped files at 2 rpm until breakage. Cutting efficiency was determined by measuring the velocity of file advancement into plastic blocks with 100-g constant force for 5 seconds. Data were analyzed by either an unpaired t test or analysis of variance followed by a Fisher probable least-square difference test. Electropolishing significantly reduced resistance to cyclic fatigue but did not affect torsional resistance. However, electropolishing reduced the angle at failure and amount of unwinding. Electropolishing did not significantly affect cutting efficiency.

Citing Articles

Revolutionizing endodontics: Advancements in nickel-titanium instrument surfaces.

Agrawal P, Chandak M, Nikhade P, Patel A, Bhopatkar J J Conserv Dent Endod. 2024; 27(2):126-133.

PMID: 38463467 PMC: 10923218. DOI: 10.4103/JCDE.JCDE_248_23.


Effect from Autoclave Sterilization and Usage on the Fracture Resistance of Heat-Treated Nickel-Titanium Rotary Files.

Abed R, Al Raeesi D, Alshehhi A, Alkhatib Z, Khamis A, Jamal M Materials (Basel). 2023; 16(6).

PMID: 36984141 PMC: 10056662. DOI: 10.3390/ma16062261.


Structural Analysis of NiTi Endodontic Instruments: A Systematic Review.

Barbosa I, Ferreira F, Scelza P, Adeodato C, Caldas I, Goncalves F Iran Endod J. 2023; 15(3):124-139.

PMID: 36703804 PMC: 9709855. DOI: 10.22037/iej.v15i3.28083.


The state of the art and future trends of root canal files from the perspective of patent analysis: a study design.

Jiang J, Sun J, Huang Z, Bi Z, Yu G, Yang J Biomed Eng Online. 2022; 21(1):90.

PMID: 36566212 PMC: 9789667. DOI: 10.1186/s12938-022-01060-0.


: From heat treatment to surface functionalization for nickel-titanium (NiTi) instruments in endodontics.

Chan W, Gulati K, Peters O Bioact Mater. 2022; 22:91-111.

PMID: 36203965 PMC: 9520078. DOI: 10.1016/j.bioactmat.2022.09.008.