Quality Assurance for Point-of-care Testing of Oral Anticoagulation: a Large-scale Evaluation of the Hemochron Junior Signature Microcoagulation System

Overview
Specialty Hematology
Date 2008 Jan 8
PMID 18177437
Citations 1
Authors
Affiliations
Soon will be listed here.
Abstract

We report the first large-scale evaluation of the Hemochron Junior Signature (HJS) Microcoagulation System for community monitoring of oral anticoagulation and establishment of a programme of internal and external quality assurance. Over 1600 HJS results, with a simultaneous venous sample for central analysis, were obtained over a 19 month period. Monitoring of an initial period of HJS results (n = 135) revealed an International Normalized Ratio (INR) over estimation (mean +1.05), with only 27% of results within 0.5 of the central laboratory INR. A correction factor was introduced which reduced the INR bias to +0.07 and improved the percentage of results within 0.5 of the central laboratory INR to 76% (n = 353). A revised correction factor was later introduced to adjust for an under estimation at higher INR values. This changed the INR bias to -0.05, with 76% of results within 0.5 of the central laboratory INR (n = 1174). Local external quality assurance samples were distributed monthly with a total of 791 samples during the study period. 84% of test results were within 15% of the median value (range 73-97% per month). These results emphasize the value of a robust quality assurance programme when using point-of-care devices for community monitoring of oral anticoagulation.

Citing Articles

Patients' perspectives on self-testing of oral anticoagulation therapy: content analysis of patients' internet blogs.

Shah S, Robinson I BMC Health Serv Res. 2011; 11:25.

PMID: 21291542 PMC: 3045880. DOI: 10.1186/1472-6963-11-25.