Pascal, ICare and Goldmann Applanation Tonometry--a Comparative Study

Overview
Journal Acta Ophthalmol
Specialty Ophthalmology
Date 2007 Dec 21
PMID 18093261
Citations 19
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To compare intraocular pressure (IOP) measurements by Pascal, ICare and Goldmann applanation tonometry (GAT), to evaluate the effects of central corneal thickness (CCT) and curvature on IOP measurement and to estimate the intra-observer variability.

Methods: A prospective, single-centre study of 150 eyes with a wide range of pressures. Six masked IOP measurements/method; corneal thickness and curvature were studied for each eye. GAT was the reference.

Results: IOPPascal and IOPICare correlated with IOPGAT (r = 0.91, 0.89). Mean ICare measurement exceeded GAT by 2 mmHg. Pascal measured higher than GAT at low IOPs and lower at high IOPs. For every 10 mmHg increase in IOP above 31 mmHg, Pascal measured 2 mmHg lower than GAT and vice versa. CCT was correlated significantly with IOPGAT (r = 0.23) and IOPICare (r = 0.43) but not with IOPPascal (P = 0.12). CCT was correlated with age. In a subgroup (>50 years), ICare and the difference between IOPGAT and IOPPascal were affected significantly by the CCT, whereas IOPGAT and IOPPascal were not. Corneal curvature was correlated significantly with IOPGAT (r = -0.27) and IOPPascal (r = -0.26) but not with IOPICare (P = 0.60). Intra-observer variability within each set of six measurements was approximately 2 mmHg, irrespective of method.

Conclusion: This study showed a reasonable overall correlation and concordance between the IOP obtained with the three instruments. None of the methods were completely independent of the biomechanical properties of the cornea. ICare showed a significant dependency upon CCT, whereas GAT and Pascal showed a significant dependency on corneal curvature. All methods showed intra-observer variability, which leaves room for further improvement of methods.

Citing Articles

A comparison of iCare and Goldmann applanation tonometry measurements during the COVID-19 pandemic: a retrospective study.

Swaminathan S, Kisielewski A, Nowroozzadeh M, Hallaj S, Lee D, Kolomeyer N Int Ophthalmol. 2024; 44(1):260.

PMID: 38910200 PMC: 11194190. DOI: 10.1007/s10792-024-03220-8.


Posture-Induced Intraocular Pressure Changes after iStent Inject W Combined with Phacoemulsification in Open Angle Glaucoma Patients.

Iwasaki K, Arimura S, Orii Y, Inatani M Medicina (Kaunas). 2023; 59(3).

PMID: 36984423 PMC: 10052109. DOI: 10.3390/medicina59030423.


Early intraocular pressure changes following different keratoplasty techniques and association with cornea parameters and anterior chamber depth.

Gumus G, Altan C, Yildirim Y, Kandemir Besek N, Genc S, Kirgiz A Ther Adv Ophthalmol. 2022; 14:25158414221083359.

PMID: 35321307 PMC: 8935592. DOI: 10.1177/25158414221083359.


Applications of Scheimpflug Imaging in Glaucoma Management: Current and Potential Applications.

Nguyen A, Liu T, Liu J J Ophthalmol. 2017; 2016:3062381.

PMID: 28044101 PMC: 5164893. DOI: 10.1155/2016/3062381.


Systematic Review of Current Devices for 24-h Intraocular Pressure Monitoring.

Ittoop S, SooHoo J, Seibold L, Mansouri K, Kahook M Adv Ther. 2016; 33(10):1679-1690.

PMID: 27531519 PMC: 5055550. DOI: 10.1007/s12325-016-0388-4.