» Articles » PMID: 18048891

The Light-tissue Interaction of Pulse Oximetry

Overview
Journal Anesth Analg
Specialty Anesthesiology
Date 2007 Dec 6
PMID 18048891
Citations 55
Authors
Affiliations
Soon will be listed here.
Abstract

The underlying science of pulse oximetry is based on a simple manipulation of the Lambert-Beer law, which describes the attenuation of light traveling through a mixture of absorbers. Signals from detected red and infrared light that has traveled through blood-perfused tissues are used to estimate the underlying arterial hemoglobin oxygen saturation. However, light scatters in tissue and influences some of the simplifications made in determining this relationship. Under most clinical circumstances, the empirical process that manufacturers use to calibrate the system during its design readily accommodates this and results in accurate readings. The same tissue light scattering properties allow sensors to be configured for use on opposing or adjacent surfaces, provided that the placement sites offer sufficient signal strength and are absent factors known to influence accuracy. In this paper I review the light-tissue interaction in pulse oximetry and describe some of the assumptions made and their implications. Certain deviations from the nominal conditions, whether clinical in nature or misuse of the product, can affect system performance. Consequently, users should be cautious in modifying sensors and/or using them on tissue sites not intended by the manufacturer (off-label use). While perhaps helpful for obtaining pulsatile signals or extending the lifetime of a sensor, some practices can disrupt the optical integrity of the measurement and negatively impact the oxygen saturation reading accuracy.

Citing Articles

MultiPhys: Heterogeneous Fusion of Mamba and Transformer for Video-Based Multi-Task Physiological Measurement.

Huo C, Yin P, Fu B Sensors (Basel). 2025; 25(1.

PMID: 39796891 PMC: 11722562. DOI: 10.3390/s25010100.


Pulse oximetry in pediatric care: Balancing advantages and limitations.

Al-Beltagi M, Saeed N, Bediwy A, Elbeltagi R World J Clin Pediatr. 2024; 13(3):96950.

PMID: 39350904 PMC: 11438930. DOI: 10.5409/wjcp.v13.i3.96950.


Raw Photoplethysmography as an Enhancement for Research-Grade Wearable Activity Monitors.

Hibbing P, Khan M JMIR Mhealth Uhealth. 2024; 12:e57158.

PMID: 39331461 PMC: 11470225. DOI: 10.2196/57158.


What Remote PPG Oximetry Tells Us about Pulsatile Volume?.

Saiko G Biomedicines. 2024; 12(8).

PMID: 39200248 PMC: 11351970. DOI: 10.3390/biomedicines12081784.


Ninety years of pulse oximetry: history, current status, and outlook.

Quaresima V, Ferrari M, Scholkmann F J Biomed Opt. 2024; 29(Suppl 3):S33307.

PMID: 39156662 PMC: 11330276. DOI: 10.1117/1.JBO.29.S3.S33307.