Proximal Humeral Fractures: What is Semi-rigid? Biomechanical Properties of Semi-rigid Implants, a Biomechanical Cadaver Based Evaluation

Overview
Date 2007 Nov 28
PMID 18040704
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: Proximal humerus fractures remain challenging especially in the elderly. Biomechanical data put semi-rigid implants in favour of osteopenic or osteoporotic situation. Little surgical side damage is associated with a minimal invasive approach of these implants. The aim of this study was to evaluate the mechanical properties of three such implants.

Material And Methods: Fresh frozen cadaver specimens were mounted as proposed by the distributors. Three different implants were used: LCP-PH (locking compression plate proximal humerus, Synthes, Austria), HB (humerus block, Synthes, Austria), and IMC (intramedullary claw, ITS, Austria). Subcapital fracture was simulated by resecting a 5 mm gap. All specimens were comparable in "B" (one), "M" (ineral) and "D" (ensity). Four load cases were tested: varus bending, medial shearing and axial torque. A cyclic test (1,000 cycles) was performed in the first load case (varus stress) for all three implants.

Results: The LCP-PH was the most rigid in all three load cases, always followed by the HB. The IMC was the most elastic device with almost immeasurable values in axial torque. In the cyclic setting, the load reduction of the HB followed by the LCP-PH was significantly better than that for the IMC.

Conclusion: The differences in stiffness are varying tremendously. The IMC is the implant with the lowest stiffness in all load cases and the highest load reduction. New "semi-rigids" claim good clinical performance, yet prospective clinical studies have to prove this. It is unlikely that the IMC can maintain fracture reduction in fracture situations of complex nature (no ligamentotaxis).

Citing Articles

Cyclic Stability of Locking Plate Augmented with Intramedullary Polymethyl Methacrylate (PMMA) Strut Fixation for Osteoporotic Humeral Fractures: A Biomechanical Study.

Hsiao C, Chiu Y, Hsiao H, Tsai Y, Lee C, Yen C Life (Basel). 2023; 13(11).

PMID: 38004250 PMC: 10672415. DOI: 10.3390/life13112110.


Anatomical variation in humeri: gender and side comparison using statistical shape modelling.

Dauwe J, Vancleef S, De Bondt S, Nijs S Int Orthop. 2023; 47(4):1013-1020.

PMID: 36763125 DOI: 10.1007/s00264-023-05713-0.


Surgical Fixation of Three- and Four-Part Proximal Humeral Fractures Using the Proximal Humeral Interlocking System Plate.

Saber A, Said U, Abdelmonem A, Elsayed H, Taha M, Hussein W Cureus. 2022; 14(5):e25348.

PMID: 35774694 PMC: 9236683. DOI: 10.7759/cureus.25348.


Biomechanical comparison of screw-based zoning of PHILOS and Fx proximal humerus plates.

Jabran A, Peach C, Zou Z, Ren L BMC Musculoskelet Disord. 2018; 19(1):253.

PMID: 30045716 PMC: 6060456. DOI: 10.1186/s12891-018-2185-5.


Biomechanical analysis of plate systems for proximal humerus fractures: a systematic literature review.

Jabran A, Peach C, Ren L Biomed Eng Online. 2018; 17(1):47.

PMID: 29703261 PMC: 5923007. DOI: 10.1186/s12938-018-0479-3.