» Articles » PMID: 18040610

Evaluation of Calculation Algorithms Implemented in Different Commercial Planning Systems on an Anthropomorphic Breast Phantom Using Film Dosimetry

Overview
Specialties Oncology
Radiology
Date 2007 Nov 28
PMID 18040610
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To evaluate the accuracy of dose calculation algorithms of different planning systems for postoperative tangential radiotherapy in breast cancer.

Material And Methods: On a CT dataset of an anthropomorphic phantom, a structure set of the left lung, clinical target volume (CTV), planning target volume, heart, and external contour were delineated. The dataset was processed by six radiation oncology centers participating in this multicenter dosimetry project. Conventional plans with two tangential wedged fields were generated in MasterPlan, Pinnacle, Eclipse, TMS, and PrecisePLAN. Plan calculations were done using the beam data of local linacs. The dose distributions were verified under local conditions with Gafchromic-EBT films.

Results: In all planning systems, deviations between calculation and measurement were around +/-3% in the CTV in the measured plane. Only small areas with deviations of +/-5% were detected. Pencil-beam (PB) calculations overestimated the dose inside the lung by up to 23%. Collapsed cone (CC) underestimated the lung dose by up to 6%.

Conclusion: CC calculates the dose distribution more accurately than PB. Inside regions with electron disequilibrium, however, the dose is slightly underestimated.

Citing Articles

Monte Carlo simulations of EBT3 film dose deposition for percentage depth dose (PDD) curve evaluation.

Robinson S, Esplen N, Wells D, Bazalova-Carter M J Appl Clin Med Phys. 2020; 21(12):314-324.

PMID: 33155768 PMC: 7769387. DOI: 10.1002/acm2.13078.


Film-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom.

Pan Y, Yang R, Li J, Zhang X, Liu L, Wang J J Appl Clin Med Phys. 2018; 19(3):142-148.

PMID: 29603564 PMC: 5978558. DOI: 10.1002/acm2.12314.


Assessment of the accuracy of dose calculation in the build-up region of the tangential field of the breast for a radiotherapy treatment planning system.

Farhood B, Toossi M, Soleymanifard S, Mortezazadeh T Contemp Oncol (Pozn). 2017; 21(3):232-239.

PMID: 29180932 PMC: 5701585. DOI: 10.5114/wo.2017.70114.


Evaluation of the uncertainty in an EBT3 film dosimetry system utilizing net optical density.

Marroquin E, Herrera Gonzalez J, Camacho Lopez M, Villarreal Barajas J, Garcia-Garduno O J Appl Clin Med Phys. 2016; 17(5):466-481.

PMID: 27685125 PMC: 5874103. DOI: 10.1120/jacmp.v17i5.6262.


Dosimetric effects of swelling or shrinking tissue during helical tomotherapy breast irradiation. A phantom study.

Klepper R, Hofel S, Botha U, Kohler P, Zwicker F J Appl Clin Med Phys. 2014; 15(4):382–391.

PMID: 25207418 PMC: 5875514. DOI: 10.1120/jacmp.v15i4.4873.