In Vitro Evaluation of Platelet Concentrates During Storage: Platelet Counts and Markers of Platelet Destruction
Overview
Authors
Affiliations
Background: Differences in platelet counts are observed by use of automated haematology analyzers making interlaboratory comparison difficult.
Materials And Methods: Twenty-eight single-donor platelet concentrates (PCs) were collected. Platelet concentration and markers of platelet destruction were investigated during storage for 11/12 days.
Results: Increasing impedance-immunoplatelet ratio was observed during storage, correlating to platelet fragments, large platelets, platelet density and cell-lysis. High variability was observed for optical-immunoplatelet ratio.
Conclusion: Immunoplatelet count or correction factor calculated by impedance-immunoplatelet ratio should be used to confirm that platelet unit meets platelet count requirements or to compare results from clinical trials. Optical platelet count is not recommended.
Hosseini E, Hojjati S, Afzalniaye Gashti S, Ghasemzadeh M Thromb J. 2020; 18:18.
PMID: 32884450 PMC: 7457792. DOI: 10.1186/s12959-020-00233-y.
Hosseini E, Beshkar P, Ghasemzadeh M J Thromb Thrombolysis. 2018; 46(4):534-540.
PMID: 30218271 DOI: 10.1007/s11239-018-1739-6.
Evaluation of platelet activation in leukocyte-depleted platelet concentrates during storage.
Vucic M, Stanojkovic Z, Antic A, Vucic J, Pavlovic V Bosn J Basic Med Sci. 2017; 18(1):29-34.
PMID: 28926321 PMC: 5826671. DOI: 10.17305/bjbms.2017.2321.
Could Microparticles Be the Universal Quality Indicator for Platelet Viability and Function?.
Maurer-Spurej E, Chipperfield K J Blood Transfus. 2017; 2016:6140239.
PMID: 28053805 PMC: 5178367. DOI: 10.1155/2016/6140239.
Mangalpally K, Siqueiros-Garcia A, Vaduganathan M, Dong J, Kleiman N, Guthikonda S J Thromb Thrombolysis. 2010; 30(3):251-62.
PMID: 20502945 DOI: 10.1007/s11239-010-0489-x.