Percutaneous Retrieval of Dislodged Totally Implantable Central Venous Access System in 92 Cases: Experience in a Single Hospital
Overview
Affiliations
Objective: To investigate the clinical presentation of dislodged totally implantable central venous access system (central venous port-catheter) fragments and the efficacy and safety of percutaneous retrieval of them in our hospital.
Materials And Methods: Ninety-two cancer patients, mean age of 53.8 years old with 51.1% male, were enrolled from January 2005 to March 2007. They were referred to our catheterization laboratory for retrieval of fractured central venous port-catheter in our hospital. All patients were followed in the outpatient department for at least 1 month after surgical insertion. The characteristics of disrupted central venous port-catheter were recorded. The procedure-related clinical condition was evaluated.
Results: The most common presentation of central venous port-catheter dislodgement is irrigation resistance to infusion (51/92). The most common location of fractured fragments is between superior vena cava and right atrium (i.e. proximal end remained in superior vena cava and distal end in right atrium) (22/92). The most common fracture site of the catheter is at the anastomosis between injection port and catheter (77/92). The retrieval set used mostly is loop snare. The success rate of the percutaneous retrieval of dislodged fragment was 97.8% and the complication rate was 3.3% only.
Conclusion: The faulty connection between catheter and injection port contributes mainly to dislodgement of central venous port-catheter. Percutaneous retrieval of dislodged catheter is a highly successful, safe and efficient method.
Dave S, Khunteta N, Mohan A, Agarwal D, Das M, Viswanath M Indian J Surg Oncol. 2024; 15(3):590-593.
PMID: 39239435 PMC: 11371993. DOI: 10.1007/s13193-024-01949-3.
Vascular Port Complication Leading to Surgery of Pulmonary Artery Branch-A Case Report.
Kordykiewicz D, Skorka P, Morozik M, Wojcik J, Wojtys M Medicina (Kaunas). 2024; 60(7).
PMID: 39064522 PMC: 11279055. DOI: 10.3390/medicina60071093.
Port Catheter Fracture: A Silent Intrusion into the Right Ventricle.
Chandna S, Nanda S, Agrawal A, Raj K, Bhagat U, Pillai K J Community Hosp Intern Med Perspect. 2024; 14(3):110-112.
PMID: 39036567 PMC: 11259478. DOI: 10.55729/2000-9666.1346.
Surgical removal of a peripheral venous catheter fragment in the heart in a preterm infant.
Wang G, Zhao Z, Zhou G, Feng Z J Cardiothorac Surg. 2024; 19(1):306.
PMID: 38822410 PMC: 11140915. DOI: 10.1186/s13019-024-02818-4.
Kamisako A, Ikoma A, Suzuki T, Kakimoto N, Suenaga T, Tokuhara D Radiol Case Rep. 2024; 19(5):1965-1969.
PMID: 38434777 PMC: 10909603. DOI: 10.1016/j.radcr.2024.02.014.