» Articles » PMID: 17968897

Interventional Cardiovascular Procedures Guided by Real-time MR Imaging: an Interactive Interface Using Multiple Slices, Adaptive Projection Modes and Live 3D Renderings

Overview
Date 2007 Oct 31
PMID 17968897
Citations 31
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To develop and test a novel interactive real-time MRI environment that facilitates image-guided cardiovascular interventions.

Materials And Methods: Color highlighting of device-mounted receiver coils, accelerated imaging of multiple slices, adaptive projection modes, live three-dimensional (3D) renderings and other interactive features were utilized to enhance navigation of devices and targeting of tissue.

Results: Images are shown from several catheter-based interventional procedures performed in swine that benefit from this custom interventional MRI interface. These include endograft repair of aortic aneurysm, balloon septostomy of the cardiac interatrial septum, angioplasty and stenting, and endomyocardial cell injection, all using active catheters containing MRI receiver coils.

Conclusion: Interactive features not available on standard clinical scanners enhance real-time MRI for guiding cardiovascular interventional procedures.

Citing Articles

Interventional device tracking under MRI via alternating current controlled inhomogeneities.

Uzun D, Yildirim D, Bruce C, Halaby R, Jaimes A, Potersnak A Magn Reson Med. 2024; 92(1):346-360.

PMID: 38394163 PMC: 11055668. DOI: 10.1002/mrm.30031.


Cardiac magnetic resonance haemodynamics in paediatric heart transplant patients: fick oximetry versus cardiac magnetic resonance phase contrast.

Schramm J, Cronin I, McCarter R, Mandell J, Downing T, Kanter J Cardiol Young. 2023; 34(2):262-267.

PMID: 37317547 PMC: 10721722. DOI: 10.1017/S1047951123001440.


An interventional MRI guidewire combining profile and tip conspicuity for catheterization at 0.55T.

Yildirim D, Uzun D, Bruce C, Khan J, Rogers T, Schenke W Magn Reson Med. 2022; 89(2):845-858.

PMID: 36198118 PMC: 9712240. DOI: 10.1002/mrm.29466.


Improved catheter tracking during cardiovascular magnetic resonance-guided cardiac catheterization using overlay visualization.

Greer J, Hussein M, Vamsee R, Arar Y, Krueger S, Weiss S J Cardiovasc Magn Reson. 2022; 24(1):32.

PMID: 35650624 PMC: 9161533. DOI: 10.1186/s12968-022-00863-3.


Heat-Mitigated Design and Lorentz Force-Based Steering of an MRI-Driven Microcatheter toward Minimally Invasive Surgery.

Phelan 3rd M, Tiryaki M, Lazovic J, Gilbert H, Sitti M Adv Sci (Weinh). 2022; 9(10):e2105352.

PMID: 35112810 PMC: 8981448. DOI: 10.1002/advs.202105352.


References
1.
Scheffler K, Heid O, Hennig J . Magnetization preparation during the steady state: fat-saturated 3D TrueFISP. Magn Reson Med. 2001; 45(6):1075-80. DOI: 10.1002/mrm.1142. View

2.
Dick A, Raman V, Raval A, Guttman M, Thompson R, Ozturk C . Invasive human magnetic resonance imaging: feasibility during revascularization in a combined XMR suite. Catheter Cardiovasc Interv. 2005; 64(3):265-74. PMC: 1317610. DOI: 10.1002/ccd.20302. View

3.
Dick A, Guttman M, Raman V, Peters D, Pessanha B, Hill J . Magnetic resonance fluoroscopy allows targeted delivery of mesenchymal stem cells to infarct borders in Swine. Circulation. 2003; 108(23):2899-904. PMC: 1325104. DOI: 10.1161/01.CIR.0000095790.28368.F9. View

4.
Feng L, Dumoulin C, Dashnaw S, Darrow R, Guhde R, DeLaPaz R . Transfemoral catheterization of carotid arteries with real-time MR imaging guidance in pigs. Radiology. 2004; 234(2):551-7. DOI: 10.1148/radiol.2341031951. View

5.
Aksit P, Derbyshire J, Serfaty J, Atalar E . Multiple field of view MR fluoroscopy. Magn Reson Med. 2002; 47(1):53-60. DOI: 10.1002/mrm.10035. View