» Articles » PMID: 17932842

A Comparison of the Polypropylene Plate Ahmed Glaucoma Valve to the Silicone Plate Ahmed Glaucoma Flexible Valve

Overview
Specialty Ophthalmology
Date 2007 Oct 13
PMID 17932842
Citations 21
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To compare surgical outcomes following implantation of the polypropylene plate Ahmed glaucoma valve (AGV) (S-2) and the newer silicone plate AGV (FP-7) with minimum 12 months follow-up in patients with intractable glaucoma.

Design: Retrospective chart review of 25 eyes of 22 patients who underwent FP-7 and 26 eyes of 22 patients who underwent S-2 AGV implantation by a single surgeon. Main outcome measures included surgical outcomes in terms of intraocular pressure (IOP) control, hypertensive phase, and postoperative complications.

Results: There were no significant differences in diagnoses, pre-and postoperative IOP, pre- and postoperative visual acuities at all time points, and in the number of pre- and postoperative medications between the groups. Significant IOP reductions occurred after FP-7 implantation (29.6 +/- 13.1 to 12.1 +/- 3.4, p < 0.001) and S-2 implantation (31.4 +/- 11.7 to 16.1 +/- 5.7, p < 0.001). The FP-7 group had a statistically lower IOP compared to the S-2 group at 6 (p<0.05), 9 (p<0.01), and 12 months (p<0.01). There was no significant difference in the hypertensive phase between the two groups (p=0.41).

Conclusions: FP-7 model silicone plate AGV resulted in lower IOP compared to the older S-2 polypropylene plate AGV at 1 year. This may suggest that the plate material may influence clinical outcome.

Citing Articles

Surgical safety and efficacy of a multi-stent system, the comet stent, in Baerveldt glaucoma implant surgery: a prospective pilot study.

Manabe S, Hirata A, Akagi T, Hayashi K Jpn J Ophthalmol. 2023; 67(6):685-692.

PMID: 37540326 DOI: 10.1007/s10384-023-01014-6.


Ahmed Glaucoma Valve Implantation to Reduce Intraocular Pressure: Updated Perspectives.

Arikan G, Gunenc U Clin Ophthalmol. 2023; 17:1833-1845.

PMID: 37405008 PMC: 10317551. DOI: 10.2147/OPTH.S342721.


Rational Design of Microfluidic Glaucoma Stent.

Graf T, Kancerevycius G, Jonusauskas L, Eberle P Micromachines (Basel). 2022; 13(6).

PMID: 35744591 PMC: 9229318. DOI: 10.3390/mi13060978.


Risk Factors for Failure of Tube Shunt Surgery: A Pooled Data Analysis.

Bowden E, Choudhury A, Gedde S, Feuer W, Christakis P, Savatovsky E Am J Ophthalmol. 2022; 240:217-224.

PMID: 35288073 PMC: 9614408. DOI: 10.1016/j.ajo.2022.02.027.


Risk factors for hypertensive phase after Ahmed glaucoma valve implantation.

Ozalp O, Ilguy S, Atalay E, Simsek T, Yildirim N Int Ophthalmol. 2021; 42(1):147-156.

PMID: 34420122 DOI: 10.1007/s10792-021-02009-3.