Abstract

Introduction: Lower radiopharmaceutical activities are used for myocardial perfusion scintigraphy (MPS) in the UK than in other countries. There is no evidence to suggest that higher activities improve imaging or clinical outcome.

Materials And Methods: We undertook a multicentre study of the relationship between radiopharmaceutical activity and imaging outcome. Fifty-one patients with clinical referrals for MPS followed a 2-day protocol with an injection of 1,000 MBq (99m)Tc-tetrofosmin for each of the stress and rest images. ECG-gated acquisition was performed in three rotations occupying 25, 35 and 40% of a standard acquisition, and rotations were summed to simulate administered activities of 250, 400, 750 and 1,000 MBq. Each set of images was reported by an experienced physician who was blinded to all clinical information and to the simulated activity. Scores were assigned for image quality, low count, attenuation and reconstruction artefact, segmental tracer uptake, segmental and global defect classification, and confidence in the global classification. The images were reported twice to assess intra-observer variability.

Results: Positive relationships were found between administered activity and overall image quality, observer confidence and intra-observer agreement of uptake score, and a negative relationship was found with low-count artefact. For the majority of comparisons, there was no additional improvement with increasing activity from 750 to 1,000 MBq. Intra-observer agreement was found to be better in female patients and in those below average body mass index.

Discussion: We conclude that higher administered radiopharmaceutical activities lead to better quality images and improved surrogate parameters for clinical outcome, but that activities above 750 MBq may be unnecessary in average patients.

Citing Articles

Performance of cardiac cadmium-zinc-telluride gamma camera imaging in coronary artery disease: a review from the cardiovascular committee of the European Association of Nuclear Medicine (EANM).

Agostini D, Marie P, Ben-Haim S, Rouzet F, Songy B, Giordano A Eur J Nucl Med Mol Imaging. 2016; 43(13):2423-2432.

PMID: 27542010 DOI: 10.1007/s00259-016-3467-5.


Comparative analysis of full-time, half-time, and quarter-time myocardial ECG-gated SPECT quantification in normal-weight and overweight patients.

Lecchi M, Martinelli I, Zoccarato O, Maioli C, Lucignani G, Del Sole A J Nucl Cardiol. 2016; 24(3):876-887.

PMID: 26911365 DOI: 10.1007/s12350-015-0382-2.


Stress-first protocol for myocardial perfusion SPECT imaging with semiconductor cameras: high diagnostic performances with significant reduction in patient radiation doses.

Perrin M, Djaballah W, Moulin F, Claudin M, Veran N, Imbert L Eur J Nucl Med Mol Imaging. 2015; 42(7):1004-11.

PMID: 25711177 DOI: 10.1007/s00259-015-3016-7.


Feasibility of SPECT myocardial perfusion imaging in the super-obese using a multi-head semiconductor camera with attenuation correction.

DeCicco A, Sokil A, Marhefka G, Reist K, Hansen C J Nucl Cardiol. 2014; 22(2):344-50.

PMID: 25373523 DOI: 10.1007/s12350-014-0018-y.


Factors affecting the myocardial activity acquired during exercise SPECT with a high-sensitivity cardiac CZT camera as compared with conventional Anger camera.

Verger A, Imbert L, Yagdigul Y, Fay R, Djaballah W, Rouzet F Eur J Nucl Med Mol Imaging. 2013; 41(3):522-8.

PMID: 24202049 DOI: 10.1007/s00259-013-2617-2.


References
1.
. Notes for guidance on the clinical administration of radiopharmaceuticals and use of sealed radioactive sources. Administration of Radioactive Substances Advisory Committee. Nucl Med Commun. 2000; 21 Suppl:S1-93. View

2.
Notghi A, Williams N, Smith N, Goyle S, Harding L . Relationship between myocardial counts and patient weight: adjusting the injected activity in myocardial perfusion scans. Nucl Med Commun. 2002; 24(1):55-9. DOI: 10.1097/00006231-200301000-00015. View

3.
Bavelaar-Croon C, Pauwels E, van der Wall E . Gated single-photon emission computed tomographic myocardial imaging: a new tool in clinical cardiology. Am Heart J. 2001; 141(3):383-90. DOI: 10.1067/mhj.2001.112780. View

4.
Germano G, Kiat H, Kavanagh P, Moriel M, Mazzanti M, Su H . Automatic quantification of ejection fraction from gated myocardial perfusion SPECT. J Nucl Med. 1995; 36(11):2138-47. View

5.
Donner A . Sample size requirements for the comparison of two or more coefficients of inter-observer agreement. Stat Med. 1998; 17(10):1157-68. DOI: 10.1002/(sici)1097-0258(19980530)17:10<1157::aid-sim792>3.0.co;2-w. View