Comparison and Reproducibility of Transthoracic Bioimpedance and Dual Beam Doppler Ultrasound Measurement of Cardiac Function in Healthy Volunteers

Overview
Specialty Pharmacology
Date 1991 Sep 1
PMID 1777364
Citations 9
Authors
Affiliations
Soon will be listed here.
Abstract

1. We compared the ease of use and reproducibility of two noninvasive methods, transthoracic electrical bioimpedance (TEB) (BoMed NCCOM3-R7) and non-imaging dual beam Doppler ultrasound (Quantascope--Vital Science), in cardiac output (CO) and stroke volume (SV) measurement in healthy volunteers at rest and during physiological stress, both short term and from day to day. 2. The TEB method was easier to use and not dependent on the operator. The TEB method was more reproducible both in the short term and from day to day. Both devices were able to detect CO and SV changes under physiological stress, but were less reproducible day to day during exercise. 3. At supine rest, the within subject coefficient of variation between time point for TEB was 4.6% for CO and 6.1% for SV, and 7.9% for CO and 7.4% for SV with Doppler. 4. The results from each device showed a linear correlation coefficient (r) of 0.69 for CO (P less than 0.0005) and 0.64 for SV (P less than 0.0005). The correlation coefficient improved to 0.76 (P less than 0.0005) when only changes in CO and SV were considered. There was no systematic difference in the changes detected by the two methods, but the individual variation was wide.

Citing Articles

Central Hemodynamics Measured During 5 Repetition Maximum Free Weight Resistance Exercise.

Howard J, McLESTER C, Evans T, McLESTER J, Calloway J Int J Exerc Sci. 2018; 11(2):342-354.

PMID: 29541330 PMC: 5841678. DOI: 10.70252/RAWV7859.


Comparative effects of ivabradine, a selective heart rate-lowering agent, and propranolol on systemic and cardiac haemodynamics at rest and during exercise.

Joannides R, Moore N, Iacob M, Compagnon P, Lerebours G, Menard J Br J Clin Pharmacol. 2006; 61(2):127-37.

PMID: 16433867 PMC: 1884997. DOI: 10.1111/j.1365-2125.2005.02544.x.


Short-term postural reflexes in diabetic patients with autonomic dysfunction.

Zhang Y, Critchley L, Tam Y, Tomlinson B Diabetologia. 2003; 47(2):304-11.

PMID: 14652721 DOI: 10.1007/s00125-003-1286-2.


Haemodynamic comparison of amlodipine and atenolol in essential hypertension using the quantascope.

Tham T, Herity N, Guy S, Silke B Br J Clin Pharmacol. 1993; 36(6):555-60.

PMID: 12959272 PMC: 1364660. DOI: 10.1111/j.1365-2125.1993.tb00414.x.


Disagreement between standard transthoracic impedance cardiography and the automated transthoracic electrical bioimpedance method in estimating the cardiovascular responses to phenylephrine and isoprenaline in healthy man.

de Mey C, Enterling D Br J Clin Pharmacol. 1993; 35(4):349-55.

PMID: 8485014 PMC: 1381544. DOI: 10.1111/j.1365-2125.1993.tb04150.x.


References
1.
Bland J, Altman D . Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986; 1(8476):307-10. View

2.
Shoemaker W, Appel P, Kram H, Nathan R, Thompson J . Multicomponent noninvasive physiologic monitoring of circulatory function. Crit Care Med. 1988; 16(5):482-90. DOI: 10.1097/00003246-198805000-00004. View

3.
Taylor S, Silke B . Is the measurement of cardiac output useful in clinical practice?. Br J Anaesth. 1988; 60(8 Suppl 1):90S-98S. DOI: 10.1093/bja/60.suppl_1.90s. View

4.
Mattar J, Baruzzi A, Diament D, Szynkier R, de Felippe Jr J, Luz P . A clinical comparison between cardiac output measured by thermodilution versus noninvasive thoracic electrical bioimpedance. Acute Care. 1986; 12(1):58-60. View

5.
Appel P, Kram H, Mackabee J, FLEMING A, Shoemaker W . Comparison of measurements of cardiac output by bioimpedance and thermodilution in severely ill surgical patients. Crit Care Med. 1986; 14(11):933-5. DOI: 10.1097/00003246-198611000-00004. View