Evaluation of a Simple, Potentially Individual Device for Exhaled Breath Temperature Measurement

Overview
Journal Respir Med
Publisher Elsevier
Specialty Pulmonary Medicine
Date 2007 Jul 13
PMID 17624752
Citations 18
Authors
Affiliations
Soon will be listed here.
Abstract

Rationale: Inflammation is a universal pathological reaction and is characterized among other things by increased heat production. The question stays whether the contribution of the inflamed lung tissues to the overall exhaled breath temperature (EBT) can be reliably detected and used in everyday clinical practice.

Methods: We have designed a simple device for assessment of EBT and explored its performance under standard indoor conditions. We made our measurements in the morning hours, documenting the ambient conditions (room temperature, humidity, atmospheric pressure), and physiological characteristics of the tested subjects (heart rate, blood pressure, otic and axillary temperature). We assessed its day-to-day reproducibility in 17 healthy volunteers and its ability to discriminate between the same subjects without respiratory disease and uncontrolled asthmatics (n=14). We also compared the EBT of the 14 asthmatics before and after anti-inflammatory treatment.

Results: No association was found between EBT and any of the ambient conditions: room temperature, atmospheric pressure and humidity. While otic and axillary temperatures, which were measured in parallel, maintained high correlation between each other (Spearman's rho=0.71, p<0.01), EBT did not show meaningful association with any of them. The EBT (degrees C) of asthmatics (median 35.45, range 34.12-36.09) was higher than that of controls (34.84, 32.29-35.84), (p=0.009, Mann-Whitney U test). Anti-inflammatory treatment brought down the EBT of the asthmatics (34.78, 33.23-36.06), (p=0.001, Wilcoxon Signed Ranks test), while significantly improving their spirometry too.

Conclusions: Measurements of EBT with the device we constructed are not significantly influenced by changes within the accepted range of a standard indoor environment. EBT represents a different characteristic of the human organism than otic and axillary temperatures. EBT is increased in uncontrolled asthmatics and decreases under anti-inflammatory treatment.

Citing Articles

Respiratory Rate Monitoring via a Fibre Bragg Grating-Embedded Respirator Mask with a Wearable Miniature Interrogator.

Limweshasin N, Castro I, Korposh S, Morgan S, Hayes-Gill B, Faghy M Sensors (Basel). 2024; 24(23).

PMID: 39686013 PMC: 11644672. DOI: 10.3390/s24237476.


Breathlessness and "exacerbation" questions predictive for incident COPD (MARKO study): data after two years of follow-up.

Vrbica Z, Steiner J, Labor M, Gudelj I, Plavec D PeerJ. 2023; 11:e16650.

PMID: 38130928 PMC: 10734450. DOI: 10.7717/peerj.16650.


Predominance of inhalation route in short-range transmission of respiratory viruses: Investigation based on computational fluid dynamics.

Chen W, Liu L, Hang J, Li Y Build Simul. 2022; 16(5):765-780.

PMID: 36575690 PMC: 9782262. DOI: 10.1007/s12273-022-0968-y.


Exhaled Breath Temperature Home Monitoring to Detect NSCLC Relapse: Results from a Pilot Study.

Carpagnano G, Popov T, Scioscia G, Ardo N, Lacedonia D, Malerba M Biomed Res Int. 2022; 2022:1515274.

PMID: 35237686 PMC: 8885209. DOI: 10.1155/2022/1515274.


Wearable Sensors for Remote Health Monitoring: Potential Applications for Early Diagnosis of Covid-19.

Mirjalali S, Peng S, Fang Z, Wang C, Wu S Adv Mater Technol. 2021; 7(1):2100545.

PMID: 34901382 PMC: 8646515. DOI: 10.1002/admt.202100545.