» Articles » PMID: 17599573

Comparison of 4 Registration Strategies for Computer-aided Maxillofacial Surgery

Overview
Publisher Wiley
Date 2007 Jun 30
PMID 17599573
Citations 18
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Surgeons have recently started to use computer-aided surgery (CAS) to assist with maxillofacial reconstructive surgery. This study evaluates four different CAS registration strategies in the maxillofacial skeleton.

Materials And Methods: Fifteen fiducial markers were placed on each of four cadaveric heads. Four registration protocols were used: 1) group 1-invasive markers, 2) group 2-skin surface, 3) group 3-bony landmark, 4) group 4-intraoral splint. Two observers registered each head twice with each of the four protocols and measured the target registration error (TRE). The process was repeated on two different navigation systems for confirmation.

Results: The mean TRE values were: invasive, 1.13 +/- 0.05 mm (P < 0.05); skin, 2.03 +/- 0.07 mm (P < 0.05); bone, 3.17 +/- 0.10 mm (P < 0.05); and splint, 3.79 +/- 0.13 mm (P < 0.05). The TRE values were consistent across CAS systems.

Conclusion: Of the techniques tested for CAS registration, invasive fiducial markers are the most accurate. Skin surface landmarks, bony landmarks, and an intraoral splint are incrementally less accurate.

Citing Articles

Comparison of the accuracy of dynamic navigation and the free hand approaches in the placement of pterygoid implants in the completely edentulous maxilla: An in vitro study.

Tao B, Wang N, Ling X, Ye L, Wu Y J Dent Sci. 2024; 19(4):2341-2347.

PMID: 39347067 PMC: 11437257. DOI: 10.1016/j.jds.2024.01.024.


Noctopus: a novel device and method for patient registration and navigation in image-guided cranial surgery.

Ozbek Y, Bardosi Z, Freysinger W Int J Comput Assist Radiol Surg. 2024; 19(12):2371-2380.

PMID: 38748051 PMC: 11607009. DOI: 10.1007/s11548-024-03135-w.


Augmented Reality-Assisted Transcanal Endoscopic Ear Surgery for Middle Ear Cholesteatoma.

Tsuchida K, Takahashi M, Nakazawa T, Kurihara S, Yamamoto K, Yamamoto Y J Clin Med. 2024; 13(6).

PMID: 38542003 PMC: 10971374. DOI: 10.3390/jcm13061780.


[Effectiveness of three-dimensional visible technique without fluoroscopy versus two-dimensional fluoroscopy in reduction of unstable pelvic fractures].

Li J, Chen H, Zhang W, Qi H, Zhu Z, Chang Z Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2023; 37(2):129-135.

PMID: 36796804 PMC: 9970777. DOI: 10.7507/1002-1892.202210073.


Reliability and accuracy of dynamic navigation for zygomatic implant placement.

Wu Y, Tao B, Lan K, Shen Y, Huang W, Wang F Clin Oral Implants Res. 2022; 33(4):362-376.

PMID: 35113463 PMC: 9305866. DOI: 10.1111/clr.13897.