Objective: The purpose of this study was to evaluate the value of ultrasound elastography (UE) in differentiating benign versus malignant lesions in the breast and compare it with conventional sonography and mammography.

Methods: From September 2004 to May 2005, 296 solid lesions from 232 consecutive patients were diagnosed as benign or malignant by mammography and sonography and further analyzed with UE. The diagnostic results were compared with histopathologic findings. The sensitivity, specificity, accuracy, positive and negative predictive values, and false-positive and -negative rates were calculated for each modality and the combination of UE and sonography.

Results: Of 296 lesions, 87 were histologically malignant, and 209 were benign. Ultrasound elastography was the most specific (95.7%) and had the lowest false-positive rate (4.3%) of the 3 modalities. The accuracy (88.2%) and positive predictive value (87.1%) of UE were higher than those of sonography (72.6% and 52.5%, respectively). The sensitivity values, negative predictive values, and false negative rates of the 3 modalities had no differences. A combination of UE and sonography had the best sensitivity (89.7%) and accuracy (93.9%) and the lowest false-negative rate (9.2%). The specificity (95.7%) and positive predictive value (89.7%) of the combination were better, and the false-positive rate (4.3%) of the combination was lower than those of mammography and sonography.

Conclusions: In a clinical trial with Chinese women, UE was superior to sonography and equal or superior to mammography in differentiating benign and malignant lesions in the breast. A combination of UE and sonography had the best results in detecting cancer and potentially could reduce unnecessary biopsy. Ultrasound elastography is a promising technique for evaluating breast lesions.

Citing Articles

Diagnostic Accuracy of Spectral Doppler Indices and Sonoelastography in Predicting Malignancy in Breast Imaging Reporting and Database System 3 Breast Lesions With Histopathology as the Reference Standard.

Prabakaran L, Aiyappan S, Ramesh S, Ramesh R, Kumaran S Cureus. 2024; 16(11):e73481.

PMID: 39677136 PMC: 11638018. DOI: 10.7759/cureus.73481.


[Reconstruction of elasticity modulus distribution base on semi-supervised neural network].

Zhang X, Peng B, Wang R, Wei X, Luo J Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2024; 41(2):262-271.

PMID: 38686406 PMC: 11058507. DOI: 10.7507/1001-5515.202306008.


Can the Clinical Utility of iBreastExam, a Novel Device, Aid in Optimizing Breast Cancer Diagnosis? A Systematic Review.

Bhimani F, Zhang J, Shah L, McEvoy M, Gupta A, Pastoriza J JCO Glob Oncol. 2023; 9:e2300149.

PMID: 38085036 PMC: 10846782. DOI: 10.1200/GO.23.00149.


A clinical-radiomics nomogram based on multimodal ultrasound for predicting the malignancy risk in solid hypoechoic breast lesions.

Shiyan G, Liqing J, Yueqiong Y, Yan Z Front Oncol. 2023; 13:1256146.

PMID: 37916158 PMC: 10616876. DOI: 10.3389/fonc.2023.1256146.


A review of bioengineering techniques applied to breast tissue: Mechanical properties, tissue engineering and finite element analysis.

Teixeira A, Martins P Front Bioeng Biotechnol. 2023; 11:1161815.

PMID: 37077233 PMC: 10106631. DOI: 10.3389/fbioe.2023.1161815.