Integrated Electrodes on a Silicon Based Ion Channel Measurement Platform
Overview
Biotechnology
Affiliations
We demonstrate the microfabrication of a low-noise silicon based device with integrated silver/silver chloride electrodes used for the measurement of single ion channel proteins. An aperture of 150 microm diameter was etched in a silicon substrate using a deep silicon reactive ion etcher and passivated with 30 nm of polytetrafluoroethylene via chemical vapor deposition. The average recorded noise in measurements of lipid bilayers was reduced by a factor of four through patterning of a 75 microm thick SU-8 layer around the aperture. Integrated electrodes were fabricated on both sides of the device and used for repeatable, stable, giga-seal bilayer formations as well as characteristic measurements of the transmembrane protein OmpF porin.
Nikoleli G, Nikolelis D, Siontorou C, Karapetis S Sensors (Basel). 2018; 18(1).
PMID: 29346326 PMC: 5796373. DOI: 10.3390/s18010284.
Artificial Lipid Membranes: Past, Present, and Future.
Siontorou C, Nikoleli G, Nikolelis D, Karapetis S Membranes (Basel). 2017; 7(3).
PMID: 28933723 PMC: 5618123. DOI: 10.3390/membranes7030038.
Ainslie K, Desai T Lab Chip. 2008; 8(11):1864-78.
PMID: 18941687 PMC: 2970504. DOI: 10.1039/b806446f.
Using ion channel-forming peptides to quantify protein-ligand interactions.
Mayer M, Semetey V, Gitlin I, Yang J, Whitesides G J Am Chem Soc. 2008; 130(4):1453-65.
PMID: 18179217 PMC: 2692494. DOI: 10.1021/ja077555f.