» Articles » PMID: 17502254

Clinical Evaluation of a Computer-aided Diagnosis (CAD) Prototype for the Detection of Pulmonary Embolism

Overview
Journal Acad Radiol
Specialty Radiology
Date 2007 May 16
PMID 17502254
Citations 22
Authors
Affiliations
Soon will be listed here.
Abstract

Rationale And Objectives: To evaluate the performance of a prototype computer-aided diagnosis (CAD) tool using artificial intelligence techniques for the detection of pulmonary embolism (PE) and the possible benefit for general radiologists.

Materials And Methods: Forty multidetector row computed tomography datasets (16/64- channel scanner) using 100 kVp, 100 mAs effective/slice, and 1-mm axial reformats in a low-frequency reconstruction kernel were evaluated. A total of 80 mL iodinated contrast material was injected at a flow rate of 5 mL/seconds. Primarily, six general radiologists marked any PE using a commercially available lung evaluation software with simultaneous, automatic processing by CAD in the background. An expert panel consisting of two chest radiologists analyzed all PE marks from the readers and CAD, also searching for additional finding primarily missed by both, forming the ground truth.

Results: The ground truth consisted of 212 emboli. Of these, 65 (31%) were centrally and 147 (69%) were peripherally located. The readers detected 157/212 emboli (74%) leading to a sensitivity of 97% (63/65) for central and 70% (103/147) for peripheral emboli with 9 false-positive findings. CAD detected 168/212 emboli (79%), reaching a sensitivity of 74% for central (48/65) and 82%(120/147) for peripheral emboli. A total of 154 CAD candidates were considered as false positives, yielding an average of 3.85 false positives/case.

Conclusions: The CAD software showed a sensitivity comparable to that of the general radiologists, but with more false positives. CAD detection of findings incremental to the radiologists suggests benefit when used as a second reader. Future versions of CAD have the potential to further increase clinical benefit by improving sensitivity and reducing false marks.

Citing Articles

Pixel-level annotated dataset of computed tomography angiography images of acute pulmonary embolism.

de Andrade J, Olescki G, Escuissato D, Oliveira L, Basso A, Salvador G Sci Data. 2023; 10(1):518.

PMID: 37542053 PMC: 10403591. DOI: 10.1038/s41597-023-02374-x.


Automated detection and segmentation of pulmonary embolisms on computed tomography pulmonary angiography (CTPA) using deep learning but without manual outlining.

Pu J, Gezer N, Ren S, Alpaydin A, Avci E, Risbano M Med Image Anal. 2023; 89:102882.

PMID: 37482032 PMC: 10528048. DOI: 10.1016/j.media.2023.102882.


IoMT-Enabled Computer-Aided Diagnosis of Pulmonary Embolism from Computed Tomography Scans Using Deep Learning.

Khan M, Shah P, Ahmad Khan I, Islam S, Ahmad Z, Khan F Sensors (Basel). 2023; 23(3).

PMID: 36772510 PMC: 9921395. DOI: 10.3390/s23031471.


Automated Detection and Characterization of Colon Cancer with Deep Convolutional Neural Networks.

Hasan M, Ali M, Rahman M, Islam M J Healthc Eng. 2023; 2022:5269913.

PMID: 36704098 PMC: 9873459. DOI: 10.1155/2022/5269913.


A deep learning approach for automated diagnosis of pulmonary embolism on computed tomographic pulmonary angiography.

Ajmera P, Kharat A, Seth J, Rathi S, Pant R, Gawali M BMC Med Imaging. 2022; 22(1):195.

PMID: 36368975 PMC: 9651891. DOI: 10.1186/s12880-022-00916-0.