Validation of the OMRON M7 (HEM-780-E) Blood Pressure Measuring Device in a Population Requiring Large Cuff Use According to the International Protocol of the European Society of Hypertension
Overview
Affiliations
Background: A high percentage of hypertensive patients present an arm circumference of over 32 cm; the use of a large cuff is therefore recommended. Validation studies are usually performed in the general population using a standard-size cuff. The aim of this study was to assess the accuracy of the Omron M7 device in a population with an arm circumference ranging from 32 to 42 cm.
Design: A validation study was performed according to the International Protocol of the European Society of Hypertension. This protocol is divided into two phases: the first phase is performed on 15 selected participants (45 pairs of blood-pressure measurements); if the device passes this phase, 18 supplementary participants are included (54 pairs of blood-pressure measurements), making a total number of 33 participants (99 pairs of blood-pressure measurements), on whom the analysis is performed.
Methods: For each participant, four blood-pressure measurements were performed simultaneously by two trained observers, using mercury sphygmomanometers fitted with a Y tube; the measurements alternated with three by the test device. The difference between the blood-pressure value given by the device and that obtained by the two observers (mean of the two observations) was calculated for each measure. The 99 pairs of blood-pressure differences were classified into three categories (<or=5, <or=10 and <or=15 mmHg). The number of differences in each category was compared with the number required by the European Society of Hypertension protocol.
Results: The Omron M7 device passed the first and the second phases of the validation process. The average differences between the two observers were 1.5+/-3.2 and -0.5+/-2.2 mmHg for systolic blood pressure and diastolic blood pressure, and those between the device and the mercury sphygmomanometer were -1.6+/-6.7 for systolic blood pressure and -0.12+/-4.0 mmHg for diastolic blood pressure Readings that differ by less than 5, 10 and 15 mmHg for systolic blood-pressure and diastolic blood-pressure values fulfill the recommendation criteria of the European Society of Hypertension protocol.
Conclusions: The Omron M7 (HEM-780-E) device fulfilled the validation criteria of the international protocol in a population with an arm circumference ranging from 32 to 42 cm.
Dalton-Alves F, Araujo M, Lucena B, Souto G, Lopes D, Lucena M BMJ Open. 2025; 14(12):e084736.
PMID: 39806596 PMC: 11664350. DOI: 10.1136/bmjopen-2024-084736.
Elbers R, de Oude K, Kastanidis T, Maes-Festen D, Oppewal A Int J Environ Res Public Health. 2022; 19(24).
PMID: 36554319 PMC: 9778210. DOI: 10.3390/ijerph192416438.
Kottke T, Anderson J, Zillhardt J, Sperl-Hillen J, OConnor P, Green B JAMA Netw Open. 2022; 5(8):e2229098.
PMID: 36044216 PMC: 9434355. DOI: 10.1001/jamanetworkopen.2022.29098.
Standing up against office sitting: A study protocol.
Gradidge P, Phaswana M, Wijndaele K, Crowther N, Draper C S Afr J Physiother. 2020; 76(1):1415.
PMID: 32935066 PMC: 7479433. DOI: 10.4102/sajp.v76i1.1415.
Arterial Stiffness Is Associated With Cytomegalovirus-Specific Senescent CD8 T Cells.
Yu H, Youn J, Kim J, Seong Y, Park S, Kim H J Am Heart Assoc. 2017; 6(9).
PMID: 28847915 PMC: 5634298. DOI: 10.1161/JAHA.117.006535.