Effects of Image Artifacts on Gray-value Density in Limited-volume Cone-beam Computerized Tomography

Overview
Publisher Elsevier
Date 2007 Apr 24
PMID 17448704
Citations 56
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: An in vitro study was designed to investigate the influence of projection data discontinuity-related artifacts in limited-volume cone-beam computerized tomography (CBCT) imaging of the jaws.

Study Design: Test objects were positioned in 4 patterns in a water-filled phantom as follows: bimandible and vertebrae, bimandible, left mandible and vertebrae, and left mandible. The CT imaging of the left molar region was performed using image intensifier (II)- and flat panel detector (FPD)-based CBCT scanners. The CT value of the mandible and the adjacent soft tissue region were analyzed for density by means of an 8-bit grayscale.

Results: The effects of artifacts were scored as the difference in relative density between the lingual and buccal soft tissue. The intensity of artifacts increased when more objects were presented outside the area being imaged. Fewer artifacts were noted in images produced by the particular FPD CBCT used in this investigation.

Conclusion: The CBCT system using an FPD resulted in fewer artifacts than the CBCT system using an II in this particular study.

Citing Articles

Improved soft-tissue visibility on cone-beam computed tomography with an image-generating artificial intelligence model using a cyclic generative adversarial network.

Fukuda M, Nozawa M, Akiyama H, Ariji E, Ariji Y Oral Radiol. 2024; 40(4):508-519.

PMID: 38941003 DOI: 10.1007/s11282-024-00763-5.


The effect of cone beam computerized tomography voxel size and the presence of root filling on the assessment of middle mesial canals in mandibular molar teeth.

Cimen T, Duzgun S, Akyuz I, Topcuoglu H Clin Oral Investig. 2024; 28(7):394.

PMID: 38916762 PMC: 11199224. DOI: 10.1007/s00784-024-05773-5.


Assessment of CBCT gray value in different regions-of-interest and fields-of-view compared to Hounsfield unit.

Yadegari A, Safi Y, Shahbazi S, Yaghoutiazar S, Ahsaie M Dentomaxillofac Radiol. 2023; 52(8):20230187.

PMID: 37874074 PMC: 10968765. DOI: 10.1259/dmfr.20230187.


The impact of CBCT-head tilting on 3D condylar segmentation reproducibility.

El Bachaoui S, Verhelst P, Vasconcelos K, Shaheen E, Coucke W, Swennen G Dentomaxillofac Radiol. 2023; 52(6):20230072.

PMID: 37606052 PMC: 10461261. DOI: 10.1259/dmfr.20230072.


Validation of bone mineral density measurement using quantitative CBCT image based on deep learning.

Park C, Kang S, Kim J, Huh K, Lee S, Heo M Sci Rep. 2023; 13(1):11921.

PMID: 37488135 PMC: 10366160. DOI: 10.1038/s41598-023-38943-8.