A Direct ROI Quantification Method for Inherent PVE Correction: Accuracy Assessment in Striatal SPECT Measurements

Overview
Date 2007 Mar 29
PMID 17390134
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: The clinical potential of striatal imaging with dopamine transporter (DAT) SPECT tracers is hampered by the limited capability to recover activity concentration ratios due to partial volume effects (PVE). We evaluated the accuracy of a least squares method that allows retrieval of activity in regions of interest directly from projections (LS-ROI).

Methods: An Alderson striatal phantom was filled with striatal to background ratios of 6:1, 9:1 and 28:1; the striatal and background ROIs were drawn on a coregistered X-ray CT of the phantom. The activity ratios of these ROIs were derived both with the LS-ROI method and with conventional SPECT EM reconstruction (EM-SPECT). Moreover, the two methods were compared in seven patients with motor symptoms who were examined with N-3-fluoropropyl-2-beta-carboxymethoxy-3-beta-(4-iodophenyl) (FP-CIT) SPECT, calculating the binding potential (BP).

Results: In the phantom study, the activity ratios obtained with EM-SPECT were 3.5, 5.3 and 17.0, respectively, whereas the LS-ROI method resulted in ratios of 6.2, 9.0 and 27.3, respectively. With the LS-ROI method, the BP in the seven patients was approximately 60% higher than with EM-SPECT; a linear correlation between the LS-ROI and the EM estimates was found (r=0.98, p=0.03).

Conclusion: The LS-ROI PVE correction capability is mainly due to the fact that the ill-conditioning of the LS-ROI approach is lower than that of the EM-SPECT one. The LS-ROI seems to be feasible and accurate in the examination of the dopaminergic system. This approach can be fruitful in monitoring of disease progression and in clinical trials of dopaminergic drugs.

Citing Articles

An energy-optimized collimator design for a CZT-based SPECT camera.

Weng F, Bagchi S, Zan Y, Huang Q, Seo Y Nucl Instrum Methods Phys Res A. 2015; 806:330-339.

PMID: 26640308 PMC: 4665111. DOI: 10.1016/j.nima.2015.09.115.


Validation of pixel-wise parametric mapping of myocardial blood flow with ¹³NH₃ PET in patients with hypertrophic cardiomyopathy.

Sciagra R, Passeri A, Cipollini F, Castagnoli H, Olivotto I, Burger C Eur J Nucl Med Mol Imaging. 2015; 42(10):1581-8.

PMID: 26121929 DOI: 10.1007/s00259-015-3101-y.


Improved regional activity quantitation in nuclear medicine using a new approach to correct for tissue partial volume and spillover effects.

Moore S, Southekal S, Park M, McQuaid S, Kijewski M, Muller S IEEE Trans Med Imaging. 2011; 31(2):405-16.

PMID: 21965196 PMC: 3741650. DOI: 10.1109/TMI.2011.2169981.


Brain metabolic correlates of dopaminergic degeneration in de novo idiopathic Parkinson's disease.

Berti V, Polito C, Ramat S, Vanzi E, De Cristofaro M, Pellicano G Eur J Nucl Med Mol Imaging. 2009; 37(3):537-44.

PMID: 19727712 DOI: 10.1007/s00259-009-1259-x.


Clinical correlation of the binding potential with 123I-FP-CIT in de novo idiopathic Parkinson's disease patients.

Berti V, Pupi A, Ramat S, Vanzi E, De Cristofaro M, Pellicano G Eur J Nucl Med Mol Imaging. 2008; 35(12):2220-6.

PMID: 18649079 DOI: 10.1007/s00259-008-0872-4.


References
1.
Chaly T, Dhawan V, Kazumata K, Antonini A, Margouleff C, DAHL J . Radiosynthesis of [18F] N-3-fluoropropyl-2-beta-carbomethoxy-3-beta-(4-iodophenyl) nortropane and the first human study with positron emission tomography. Nucl Med Biol. 1996; 23(8):999-1004. DOI: 10.1016/s0969-8051(96)00155-2. View

2.
Cooke B, Evans A . A phantom to assess quantitative recovery of positron tomographs. J Comput Assist Tomogr. 1983; 7(5):876-80. DOI: 10.1097/00004728-198310000-00023. View

3.
Minoshima S, Koeppe R, Frey K, Kuhl D . Anatomic standardization: linear scaling and nonlinear warping of functional brain images. J Nucl Med. 1994; 35(9):1528-37. View

4.
Soret M, Koulibaly P, Darcourt J, Hapdey S, Buvat I . Quantitative accuracy of dopaminergic neurotransmission imaging with (123)I SPECT. J Nucl Med. 2003; 44(7):1184-93. View

5.
Huesman R . A new fast algorithm for the evaluation of regions of interest and statistical uncertainty in computed tomography. Phys Med Biol. 1984; 29(5):543-52. DOI: 10.1088/0031-9155/29/5/007. View