» Articles » PMID: 1737527

Accuracy of a Filter Paper Method for Measuring Glycosylated Hemoglobin

Overview
Journal Diabetes Care
Specialty Endocrinology
Date 1992 Jan 1
PMID 1737527
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To compare glycosylated hemoglobin (GH) results obtained by filter paper fingerstick collection and mailed for assay by affinity chromatography with results from a venous sample assayed by ion-exchange chromatography (HbA1) in a local laboratory.

Research Design And Methods: Fifty-eight volunteer subjects with insulin-dependent diabetes mellitus (IDDM), aged 5-24 yr, included patients at a referral-based IDDM clinic and subjects in an ongoing research study. We obtained two blood samples from each subject. One was collected by fingerstick onto filter paper, the other by venipuncture into a vacutainer. We sent filter paper samples to the Diabetes Research Laboratory (Univ. of Missouri, Columbia, MO) for analysis. Vacutainer samples were sent to the Clinical Chemistry Department of the Clinical Laboratory, University of Colorado Health Sciences Center.

Results: Results were highly correlated (r = 0.89, P = 0.0001). Fifty-nine percent were classified identically when results were normalized to SD units and grouped to suggest levels of clinical concern.

Conclusions: The filter paper method is a convenient, accurate measure of glycosylated hemoglobin in young people with IDDM. It should be considered a credible alternative research and clinical tool.

Citing Articles

HbA determination from HemaSpot™ blood collection devices: comparison of home prepared dried blood spots with standard venous blood analysis.

Hall J, Fowler C, Barrett F, Humphry R, Van Drimmelen M, MacRury S Diabet Med. 2019; 37(9):1463-1470.

PMID: 31418916 PMC: 7496699. DOI: 10.1111/dme.14110.


The use of dried blood spot sampling for the measurement of HbA1c: a cross-sectional study.

Mastronardi C, Whittle B, Tunningley R, Neeman T, Paz-Filho G BMC Clin Pathol. 2015; 15:13.

PMID: 26157353 PMC: 4495815. DOI: 10.1186/s12907-015-0013-5.


Comparability of HbA1c and lipids measured with dried blood spot versus venous samples: a systematic review and meta-analysis.

Affan E, Praveen D, Chow C, Neal B BMC Clin Pathol. 2014; 14:21.

PMID: 25045323 PMC: 4101836. DOI: 10.1186/1472-6890-14-21.


Results of a successful telephonic intervention to improve diabetes control in urban adults: a randomized trial.

Walker E, Shmukler C, Ullman R, Blanco E, Scollan-Koliopoulus M, Cohen H Diabetes Care. 2011; 34(1):2-7.

PMID: 21193619 PMC: 3005454. DOI: 10.2337/dc10-1005.


The use of the dried blood spot sample in epidemiological studies.

Parker S, Cubitt W J Clin Pathol. 2000; 52(9):633-9.