Evaluation of a Portable Lactate Analyzer (Lactate Scout) in Dogs

Overview
Journal Vet Clin Pathol
Date 2007 Feb 22
PMID 17311192
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Measurement of blood lactate concentration has become a common practice in canine medicine. However, the accuracy of portable lactate monitors has not been reported in dogs.

Objectives: The aim of this study was to evaluate the accuracy and precision of a portable analyzer (Lactate-Scout) in measuring canine blood lactate concentration.

Methods: A preliminary study was performed to assess the effects of sample storage time and temperature on plasma lactate concentration. Blood samples obtained from 6 canine patients at our hospital were divided into 8 aliquots and stored at 4 degrees C and 20 degrees C; plasma lactate was measured in duplicate with a spectrophotometric system (Konelab) at 0, 30, 60, 120, and 240 minutes after the blood collection. Values were compared with those obtained immediately after blood collection. Lactate values obtained by the portable method also were compared with those obtained by the reference spectrophotometric analyzer on blood samples collected from 48 additional canine patients.

Results: There was no significant effect of storage time (P = .89) or temperature (P = .51) on plasma lactate levels. The correlation between lactate values measured with the Lactate-Scout and the Konelab method was r = .98 (slope = .81, 95% confidence interval = .73-.87; intercept = .20, 95% confidence interval = .13-.31). The level of agreement between the 2 methods was generally good for mean lactate concentrations <5 mmol/L. However, at higher lactate concentrations (5 of 48 samples), the values recorded by the Lactate-Scout analyzer were lower than those measured by the Konelab method.

Conclusion: The Lactate-Scout analyzer is reliably comparable to a reference method for measuring whole blood lactate concentration in dogs; however, caution should be used when interpreting lactate values of 5 mmol/L and higher.

Citing Articles

Comparison of lactate measurements from earlobe and fingertip capillary blood using Biosen S-Line and lactate scout analyzers.

Zhong F, Chen Z, Gu Z, Wang X, Holmberg H, Li Y Eur J Appl Physiol. 2024; 125(1):145-156.

PMID: 39164441 DOI: 10.1007/s00421-024-05585-1.


Proposed Protocol for Field Testing of Endurance Fitness of Young Labrador Retrievers.

Soderlund E, Kyrolainen H, Laitinen-Vapaavuori O, Hyytiainen H Methods Protoc. 2023; 6(4).

PMID: 37489428 PMC: 10366876. DOI: 10.3390/mps6040061.


Correlation between L-Lactate Concentrations in Beef Cattle, Obtained Using a Hand-Held Lactate Analyzer and a Lactate Assay Colorimetric Kit.

Melendez D, Marti S, Faucitano L, Haley D, Schwinghamer T, Schwartzkopf-Genswein K Animals (Basel). 2021; 11(4).

PMID: 33806010 PMC: 8064460. DOI: 10.3390/ani11040926.


Evaluation of a handheld point-of-care analyser for measurement of creatinine in cats.

Reeve J, Warman S, Lewis D, Watson N, Papasouliotis K J Feline Med Surg. 2015; 19(2):207-215.

PMID: 26701957 PMC: 10816565. DOI: 10.1177/1098612X15622676.


Reliability of an incremental exercise test to evaluate acute blood lactate, heart rate and body temperature responses in Labrador retrievers.

Ferasin L, Marcora S J Comp Physiol B. 2009; 179(7):839-45.

PMID: 19455341 DOI: 10.1007/s00360-009-0367-z.