» Articles » PMID: 17296710

Vena Cava Filter Performance Based on Hemodynamics and Reported Thrombosis and Pulmonary Embolism Patterns

Overview
Date 2007 Feb 14
PMID 17296710
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Three inferior vena cava (IVC) filters of different designs were studied to identify the potential links between published clinical results for thrombosis and recurrent pulmonary embolism (PE) rates and in vitro hemodynamics patterns in the region of the filters.

Materials And Methods: The filters studied were the Greenfield over-the-wire filter (Medi-tech/Boston Scientific, Watertown, Mass), TrapEase filter (Cordis Europa, Roden, the Netherlands), and Mobin-Uddin umbrella filter (Edwards Laboratories, Santa Ana, Calif). To assess hemodynamics, velocity contour maps were generated for each filter by using the in vitro photochromic flow visualization technique. Results were obtained for both the unoccluded and partially occluded states. Steady flow (R(e) = 600) was used to model physiologic conditions. To estimate the rates of IVC occlusion and recurrent PE, the authors analyzed published clinical studies spanning more than 30 years and a U.S. Food and Drug Administration database.

Results: For both the unoccluded and partially occluded Mobin-Uddin and TrapEase filters, regions of flow stagnation and/or recirculation and turbulence developed downstream of the filter. The Greenfield filter did not produce any prothrombotic flow patterns for either the unoccluded or partially occluded states. Results of published clinical studies supported the hemodynamic findings, with the TrapEase and Mobin-Uddin filters having high rates of IVC occlusion and recurrent PE compared with those of the Greenfield filter.

Conclusions: Flow stagnation or recirculation and turbulence have been linked to thrombosis and thrombus and/or PE formation. Thus, the hemodynamic results from this study may help explain the relatively higher rates of filter thrombosis and PE for the Mobin-Uddin and TrapEase filters versus the Greenfield filter.

Citing Articles

Impact of Vein Wall Hyperelasticity and Blood Flow Turbulence on Hemodynamic Parameters in the Inferior Vena Cava with a Filter.

Moradicheghamahi J, Goswami D Micromachines (Basel). 2025; 16(1).

PMID: 39858706 PMC: 11767820. DOI: 10.3390/mi16010051.


Percutaneous retrieval of a Mobin-Uddin inferior vena cava filter 42 years following implantation.

Patel R, Ahuja A, Resnick S Diagn Interv Radiol. 2021; 27(6):792-795.

PMID: 34792036 PMC: 8621636. DOI: 10.5152/dir.2021.20984.


Hemodynamic Analysis of VenaTech Convertible Vena Cava Filter Using Computational Fluid Dynamics.

Wang J, Huang W, Zhou Y, Han F, Ke D, Lee C Front Bioeng Biotechnol. 2020; 8:556110.

PMID: 33195121 PMC: 7661937. DOI: 10.3389/fbioe.2020.556110.


Inferior Vena Cava Filter-Related Thrombus/Deep Vein Thrombosis: Data and Management.

Andreoli J, Thornburg B, Hickey R Semin Intervent Radiol. 2016; 33(2):101-4.

PMID: 27247478 PMC: 4862844. DOI: 10.1055/s-0036-1581087.


Error by distraction.

Darcy M Semin Intervent Radiol. 2013; 29(4):337-40.

PMID: 24293808 PMC: 3577626. DOI: 10.1055/s-0032-1330069.