Prediction of Aortic Peak Enhancement in Monophasic Contrast Injection Protocols at Multidetector CT: Phantom and Patient Studies

Overview
Journal Radiat Med
Date 2007 Jan 17
PMID 17225048
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: The aim of this study was to investigate whether it is possible to predict aortic peak enhancement (APE) from the contrast dose and injection rate.

Materials And Methods: We first undertook an experimental study using a flow phantom that simulates the human circulation. We delivered 90-150 ml of iomeprol-350 at various injection rates and measured the APE values of the simulated aorta. In our clinical study we randomized 20 patients into four groups. In groups A, B, and C the iodine dose per kilogram of body weight (BW) ranged from 450 to 600 mg, and the injection duration was fixed at 30 s; group D received 450 mg/kg over 25 s. We then measured APE in all patients at the whole aorta, averaged the three highest values, and took the result as APE.

Results: In the phantom study, the decision coefficient for the best-fit equation obtained by multiple regression analysis of the relation between the iodine dose and injection rate and the simulated APE was high (0.93). In the patient study, the predicted APE values almost corresponded with the averaged APE values when we applied the fitness equation.

Conclusion: Using our fitness equation, APE on contrast-enhanced computed tomography can be predicted from the iodine dose and the contrast injection rate per patient weight.

Citing Articles

Does Contrast Dose Based in Lean body Weight Allow Lesser Volumes on High BMI Patients for CT Angiography?.

Chandrasekharan R, Kulkarni C, Pullara S, Moorthy S J Clin Imaging Sci. 2021; 11:38.

PMID: 34345528 PMC: 8326097. DOI: 10.25259/JCIS_97_2021.


The influence of cardiac function on coronary arterial enhancement at coronary computed tomography angiography: A cross-sectional study.

Moradi M, Hashemi P, Momeni M J Res Med Sci. 2017; 21:132.

PMID: 28331518 PMC: 5348831. DOI: 10.4103/1735-1995.196614.


The Effect of Contrast Material on Radiation Dose at CT: Part I. Incorporation of Contrast Material Dynamics in Anthropomorphic Phantoms.

Sahbaee P, Segars W, Marin D, Nelson R, Samei E Radiology. 2017; 283(3):739-748.

PMID: 28092496 PMC: 5452884. DOI: 10.1148/radiol.2016152851.


Outcomes of Endovascular Aortic Aneurysm Repair in Kidney Transplant Recipients: Results From a National Quality Initiative.

Bostock I, Zarkowsky D, Hicks C, Stone D, Eslami M, Malas M Am J Transplant. 2016; 16(8):2395-400.

PMID: 26813253 PMC: 5292261. DOI: 10.1111/ajt.13733.


Low-dose contrast protocol using the test bolus technique for 64-detector computed tomography coronary angiography.

Nakaura T, Awai K, Yanaga Y, Namimoto T, Utsunomiya D, Hirai T Jpn J Radiol. 2011; 29(7):457-65.

PMID: 21882087 DOI: 10.1007/s11604-011-0579-5.

References
1.
Bae K . Peak contrast enhancement in CT and MR angiography: when does it occur and why? Pharmacokinetic study in a porcine model. Radiology. 2003; 227(3):809-16. DOI: 10.1148/radiol.2273020102. View

2.
Prokop M . Multislice CT angiography. Eur J Radiol. 2000; 36(2):86-96. DOI: 10.1016/s0720-048x(00)00271-0. View

3.
Sheiman R, Raptopoulos V, Caruso P, Vrachliotis T, PEARLMAN J . Comparison of tailored and empiric scan delays for CT angiography of the abdomen. AJR Am J Roentgenol. 1996; 167(3):725-9. DOI: 10.2214/ajr.167.3.8751690. View

4.
Becker C, Hong C, Knez A, Leber A, Bruening R, Schoepf U . Optimal contrast application for cardiac 4-detector-row computed tomography. Invest Radiol. 2003; 38(11):690-4. DOI: 10.1097/01.rli.0000084886.44676.e4. View

5.
Awai K, Hiraishi K, Hori S . Effect of contrast material injection duration and rate on aortic peak time and peak enhancement at dynamic CT involving injection protocol with dose tailored to patient weight. Radiology. 2003; 230(1):142-50. DOI: 10.1148/radiol.2301021008. View