» Articles » PMID: 17218178

Intra-articular Pressure Measurement in the Radioulnocarpal Joint Using a Novel Sensor: in Vitro and in Vivo Results

Overview
Journal J Hand Surg Am
Publisher Elsevier
Date 2007 Jan 16
PMID 17218178
Citations 22
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To determine the amount and distribution of forces transmitted across the human radioulnocarpal joint under physiologic conditions in vivo. We performed an in vitro validation of a specifically developed capacitive pressure-sensor device and an in vivo measurement of force transmission and pressure distribution at the radioulnocarpal joint in a healthy volunteer.

Methods: A new capacitive pressure-sensor device was adapted for intra-articular pressure measurement in the human radioulnocarpal joint. The technical characteristics of the device were determined and the sensor was validated in fresh-frozen cadaver tests. Force transmission across the radioulnocarpal joint then was measured in healthy volunteers under local anesthesia.

Results: The sensor delivered reproducible measurements of forces across the radioulnocarpal joint and their distribution in the cadaver experiment. In vivo, 2 centers of force transmission were identified. None of these centers correlated with previous findings in the literature. More force is transmitted across the ulnar side of the radioulnocarpal joint than previously thought. The results are consistent with clinical findings.

Conclusions: The novel sensor device is suitable for intra-articular pressure measurement in the human radioulnocarpal joint in vitro and in vivo.

Citing Articles

Establishment and preliminary evaluation of CT-based classification for distal radius fracture.

Zhang J, Yao X, Song Y, Yin P Sci Rep. 2024; 14(1):9673.

PMID: 38671052 PMC: 11053106. DOI: 10.1038/s41598-024-60416-9.


Volar fixed plating of distal radius fractures: optimizing plate position for enhanced clinical outcomes.

Emet A, Veizi E, Karaman Y, Akgun E, Tolunay T, Firat A BMC Musculoskelet Disord. 2024; 25(1):320.

PMID: 38654260 PMC: 11036783. DOI: 10.1186/s12891-024-07415-z.


Management of isolated volar lunate facet fractures of the distal radius.

Tuano K, Mudgal C J Orthop. 2024; 51:46-53.

PMID: 38299064 PMC: 10827500. DOI: 10.1016/j.jor.2024.01.002.


Combined approach versus single Henry approach for fixation of die-punch distal radius fractures: a retrospective study.

Xu Z, Liang Y, Geng G, Mu W, Xu P BMC Surg. 2023; 23(1):172.

PMID: 37355595 PMC: 10290794. DOI: 10.1186/s12893-023-02047-x.


Intra-Articular Antegrade Intramedullary Screw Fixation for Proximal Phalanx Fractures: Impact of Articular Surface Defects on Joint Contact Pressures.

Boe C, Telfer S, Kennedy S Hand (N Y). 2023; 19(8):1308-1313.

PMID: 37232109 PMC: 11536710. DOI: 10.1177/15589447231174045.