Differences in the Mechanical Properties of Calcaneal Artificial Specimens, Fresh Frozen Specimens, and Embalmed Specimens in Experimental Testing

Overview
Journal Foot Ankle Int
Publisher Sage Publications
Specialty Orthopedics
Date 2007 Jan 9
PMID 17207443
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Artificial calcanei, fresh-frozen cadaver specimens, and embalmed cadaver specimens were compared in experimental testing under biocompatible loading to clarify the biocompatibility of artificial calcaneal specimens for implant testing.

Methods: Two different artificial calcaneal bone models (Sawbone, Pacific Research Laboratories, Vashon, WA, and Synbone, Synbone Inc., Davos, Switzerland), embalmed cadaver calcaneal specimens (bone density, 313.1 +/- 40.9 g/cm2; age, 43.8 +/- 7.9 years), and fresh-frozen cadaver calcanei (bone density, 238.5 +/- 30.0 g/cm2; age, 44.4 +/- 8.2 years) were used for testing. Seven specimens of each model or cadaver type were tested. A mechanical testing machine (Zwick Inc., Ulm, Germany) was used for loading and measurements. Cyclic loading (preload 20 N, load was increased every 100 cycles by 100 N from 1,000 to 2,500 N, 0.5 mm/s) and load to failure (0.5 mm/s) were performed. The loads were applied through an artificial talus in a physiological loading direction. The displacement of the posterior facet in the primary loading direction was measured.

Results: The four different specimen groups showed different stability and different displacement in the primary loading direction during cyclic loading. The variation of the maximal displacement in the primary loading direction for the entire cyclic loading was higher in artificial specimens than in the cadaver specimens.

Conclusions: Artificial calcanei (Sawbone, Synbone) showed different biomechanical characteristics than cadaver bones (embalmed and fresh-frozen) in this experimental setup with biocompatible cyclic loading. These results do not support the use of artificial calcanei for biomechanical implant testing. Fresh-frozen and embalmed specimens seem to be equally adequate for mechanical testing. The low variation of mechanical strength in the unpaired cadaver specimens suggests that the use of PAIRED specimens is not necessary.

Citing Articles

Compressive and tensile properties of polyurethane foam mimicking trabecular tissue in artificial femoral head bones.

Zerdzicki K, Znaczko A, Kondrusik A, Korbut W Front Bioeng Biotechnol. 2025; 12:1482165.

PMID: 39898277 PMC: 11782283. DOI: 10.3389/fbioe.2024.1482165.


Elastic properties of 3D printed clavicles are closer to cadaveric bones of elderly donors than commercial synthetic bones.

Strand K, Silvestro E, Naqvi I, Hast M J Mech Behav Biomed Mater. 2024; 160:106774.

PMID: 39413544 PMC: 11560652. DOI: 10.1016/j.jmbbm.2024.106774.


Characterization of the mechanical properties of human parietal bones preserved in modified larssen solution, formalin and as fresh frozen.

Pekedis M, Yoruk M, Binboga E, Yildiz H, Bilge O, Celik S Surg Radiol Anat. 2021; 43(12):1933-1943.

PMID: 33954823 DOI: 10.1007/s00276-021-02762-1.


Material characterization and selection for 3D-printed spine models.

Hao J, Nangunoori R, Wu Y, Rajaraman M, Cook D, Yu A 3D Print Med. 2019; 4(1):8.

PMID: 30649649 PMC: 6195498. DOI: 10.1186/s41205-018-0032-9.


Integrity of a Single Superior Border Plate Repair in Mandibular Angle Fracture: A Novel Cadaveric Human Mandible Model.

Sittitavornwong S, Denson D, Ashley D, Walma D, Potter S, Freind J J Oral Maxillofac Surg. 2018; 76(12):2611.e1-2611.e8.

PMID: 30172764 PMC: 6289696. DOI: 10.1016/j.joms.2018.07.029.