» Articles » PMID: 17185663

Multi-detector Row CT Attenuation Measurements: Assessment of Intra- and Interscanner Variability with an Anthropomorphic Body CT Phantom

Overview
Journal Radiology
Specialty Radiology
Date 2006 Dec 23
PMID 17185663
Citations 85
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To determine the dependence of absolute computed tomographic (CT) attenuation values on multi-detector row CT scanner type, convolution kernel, and tube current by using an anthropomorphic phantom.

Materials And Methods: A customized phantom was designed with tissue-equivalent materials to simulate contrast material-enhanced liver, spleen, pancreas, aorta, kidney, 0- and 50-HU cylindric renal cysts, muscle, and fat. The phantom was scanned with five multi-detector row CT scanners (LightSpeed QXi, GE Healthcare, Milwaukee, Wis; MX8000, Philips Medical Systems, Best, the Netherlands; and Volume Zoom, Sensation 16 and Sensation 64, Siemens Medical Solutions, Forchheim, Germany) on five separate occasions with 120 kVp, low and high tube current settings, 3.00-3.75-mm section thickness, 50% overlap, and standard and high-spatial-resolution kernels. Standardized regions of interest (ROIs) were used to obtain 3510 attenuation measurements. Attenuation dependence on scanner, kernel, and tube current was evaluated by using F tests derived with mixed-model regression. Within the mixed-model framework, the Tukey honestly significant difference procedure and a Bonferroni multiple comparison correction were used to assess differences among imaging regimens and tube current settings, respectively, in terms of tissue attenuation and ROI standard deviation.

Results: Tube current had no significant effect (P>.4) on observed tissue attenuation. Significant (P<.0001) differences were observed between imaging regimens with respect to mean attenuation for each tissue type. Convolution kernel modification had an inconsistent effect on tissue attenuation, depending on the scanner. All multi-detector row CT scanners displayed intrascanner variability in tissue attenuation (minimum range: 8.4 HU for fat tissue with the Sensation 16; maximum range: 63.4 HU for liver tissue with the Sensation 64). The scanners behaved differently at the lower range of the CT number scale, where 0-HU cyst attenuation ranged from -15.7 to 23.9 HU and one vendor's equipment showed significantly lower mean attenuation values.

Conclusion: CT attenuation values vary significantly between different manufacturers' multi-detector row CT scanners, among different generations of multi-detector row CT scanning equipment, and with individual combinations of scanner and convolution kernel.

Supplemental Material: http://radiology.rsnajnls.org/cgi/content/full/242/1/109/DC1

Citing Articles

Super-resolution deep learning reconstruction for improved quality of myocardial CT late enhancement.

Takafuji M, Kitagawa K, Mizutani S, Hamaguchi A, Kisou R, Sasaki K Jpn J Radiol. 2025; .

PMID: 40072715 DOI: 10.1007/s11604-025-01760-2.


Intermuscular adipose tissue affected muscle density more than intramuscular adipose tissue content with opportunistic screening at abdominal CT.

Guo X, Cao N, Deng X, Wang N, Li R, Ren S Sci Rep. 2025; 15(1):8172.

PMID: 40059242 PMC: 11891326. DOI: 10.1038/s41598-025-85946-8.


Performance of different CT enhancement quantification methods as predictors of pancreatic cancer recurrence after upfront surgery.

Mohamed S, Barlemann A, Steinle V, Nonnenmacher T, Guttlein M, Hackert T Sci Rep. 2024; 14(1):19783.

PMID: 39187515 PMC: 11347575. DOI: 10.1038/s41598-024-70441-3.


Stroke recurrence and osteoporotic conditions in postmenopausal patients with atherosclerotic ischemic stroke.

Ha B, Kang S, Choi B, Cheong J, Ryu J, Won Y Heliyon. 2024; 10(9):e30196.

PMID: 38720741 PMC: 11076908. DOI: 10.1016/j.heliyon.2024.e30196.


Effect of deep learning image reconstruction with high-definition standard scan mode on image quality of coronary stents and arteries.

Liu M, Chen X, Liu W, Guo Y, Zhu Y, Duan Y Quant Imaging Med Surg. 2024; 14(2):1616-1635.

PMID: 38415168 PMC: 10895123. DOI: 10.21037/qims-23-1064.