Dosimetric Feasibility of Cone-beam CT-based Treatment Planning Compared to CT-based Treatment Planning

Overview
Specialties Oncology
Radiology
Date 2006 Oct 24
PMID 17056197
Citations 83
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Cone-beam computed tomography (CBCT) images are currently used for positioning verification. However, it is yet unknown whether CBCT could be used in dose calculation for replanning in adaptive radiation therapy. This study investigates the dosimetric feasibility of CBCT-based treatment planning.

Methods And Materials: Hounsfield unit (HU) values and profiles of Catphan, homogeneous/inhomogeneous phantoms, and various tissue regions of patients in CBCT images were compared to those in CT. The dosimetric consequence of the HU variation was investigated by comparing CBCT-based treatment plans to conventional CT-based plans for both phantoms and patients.

Results: The maximum HU difference between CBCT and CT of Catphan was 34 HU in the Teflon. The differences in other materials were less than 10 HU. The profiles for the homogeneous phantoms in CBCT displayed reduced HU values up to 150 HU in the peripheral regions compared to those in CT. The scatter and artifacts in CBCT became severe surrounding inhomogeneous tissues with reduced HU values up to 200 HU. The MU/cGy differences were less than 1% for most phantom cases. The isodose distributions between CBCT-based and CT-based plans agreed very well. However, the discrepancy was larger when CBCT was scanned without a bowtie filter than with bowtie filter. Also, up to 3% dosimetric error was observed in the plans for the inhomogeneous phantom. In the patient studies, the discrepancies of isodose lines between CT-based and CBCT-based plans, both 3D and IMRT, were less than 2 mm. Again, larger discrepancy occurred for the lung cancer patients.

Conclusion: This study demonstrated the feasibility of CBCT-based treatment planning. CBCT-based treatment plans were dosimetrically comparable to CT-based treatment plans. Dosimetric data in the inhomogeneous tissue regions should be carefully validated.

Citing Articles

A 4D-CBCT correction network based on contrastive learning for dose calculation in lung cancer.

Cao N, Wang Z, Ding J, Zhang H, Zhang S, Gao L Radiat Oncol. 2024; 19(1):20.

PMID: 38336759 PMC: 11321211. DOI: 10.1186/s13014-024-02411-y.


Does a patient-specific bladder-filling protocol affect bladder volume and dose in postprostatectomy radiotherapy?.

Yalman D, Koylu M, Duran O Int Urol Nephrol. 2023; 56(3):999-1006.

PMID: 37821762 DOI: 10.1007/s11255-023-03814-8.


Investigation of Computed Tomography Numbers on Multiple Imaging Systems using Single and Multislice Methods.

Mehta A, Arrington D, Ramachandran P, Motley R, Seshadri V, Anderson D J Med Phys. 2023; 48(1):26-37.

PMID: 37342607 PMC: 10277306. DOI: 10.4103/jmp.jmp_3_23.


Dosimetric variation in preoperative partial breast radiosurgery assessed by deformable image registrations.

Yoo S, Blitzblau R, McDuff S, Yin F, Cui Y J Radiosurg SBRT. 2023; 8(3):227-235.

PMID: 36861003 PMC: 9970744.


Dosimetric comparison of deformable image registration and synthetic CT generation based on CBCT images for organs at risk in cervical cancer radiotherapy.

Chang Y, Liang Y, Yang B, Qiu J, Pei X, Xu X Radiat Oncol. 2023; 18(1):3.

PMID: 36604687 PMC: 9817400. DOI: 10.1186/s13014-022-02191-3.