Comparison of the Innofluor Certican Assay with HPLC-UV for the Determination of Everolimus Concentrations in Heart Transplantation

Overview
Journal Clin Biochem
Specialty Biochemistry
Date 2006 Oct 24
PMID 17054930
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Objectives: Therapeutic monitoring of everolimus with chromatographic methods (HPLC) enabled effective immunosuppression while limiting the incidence of drug-related adverse events. A fluorescence polarization immunoassay (FPIA) has been recently developed for the assessment of everolimus levels. The present study was designed to evaluate FPIA performance and to compare it to HPLC.

Design And Methods: The performance of HPLC and FPIA was initially tested using drug-free whole blood spiked with different amount of everolimus concentrations and, subsequently, by analyzing 113 trough blood samples from heart transplant recipients chronically given everolimus as part of their immunosuppressive regimen.

Results: Inaccuracy and imprecision of both methods were below 15%. The correlation between everolimus concentrations and measured FPIA and HPLC was good, with a Pearson coefficient of 0.9118. The FPIA gave a mean overestimation of 24.3% as compared with HPLC. As additional analysis, cross-reactivity ranging from 85.4% to 138.0% was found with sirolimus, an immunosuppressant with a chemical structure close to everolimus.

Conclusion: The FPIA demonstrated acceptable performance for therapeutic drug monitoring of everolimus, and is a viable alternative to HPLC-based methods.

Citing Articles

Comparison of LC-MS/MS and EMIT methods for the precise determination of blood sirolimus in children with vascular anomalies.

Zhao Y, Dai H, Li Y, Zhang Y, Guo H, Ding X Front Pharmacol. 2022; 13:925018.

PMID: 36147342 PMC: 9486013. DOI: 10.3389/fphar.2022.925018.


Determination of blood sirolimus concentrations in liver and kidney transplant recipients using the Innofluor fluorescence polarization immunoassay: comparison with the microparticle enzyme immunoassay and high-performance liquid....

Bouzas L, Hermida J, Tutor J Ups J Med Sci. 2009; 114(1):55-61.

PMID: 19242874 PMC: 2852751. DOI: 10.1080/03009730802608254.


Comparison between a liquid chromatography-tandem mass spectrometry assay and a fluorescent polarization immunoassay to measure whole blood everolimus concentration in heart and renal transplantations.

Dailly E, Deslandes G, Hourmant M, Petit T, Renaud C, Treilhaud M J Clin Lab Anal. 2008; 22(4):282-5.

PMID: 18623123 PMC: 6649002. DOI: 10.1002/jcla.20258.