Assessment of Coronary Artery Stent Restenosis by 64-slice Multi-detector Computed Tomography

Overview
Journal Eur Heart J
Date 2006 Oct 13
PMID 17035252
Citations 44
Authors
Affiliations
Soon will be listed here.
Abstract

Aims: We investigated the feasibility of assessing coronary artery stent restenosis using a new generation 64-slice multi-detector computed tomography-scanner (MDCT) in comparison to conventional quantitative angiography.

Methods And Results: MDCT was performed in 64 consecutive patients (mean age 58+/-10 years) with previously implanted coronary artery stents (102 stented lesions: mean stent diameter 3.17+/-0.38 mm). Each stent was classified as 'evaluable' or 'unevaluable', and in evaluable stents, the presence of in-stent restenosis (diameter reduction >50%) was determined visually. Results were verified against invasive, quantitative coronary angiography. Fifty-nine stented lesions (58%) were classified as evaluable in MDCT. The mean diameter of evaluable stents was 3.28+/-0.40 mm, whereas the mean diameter of non-evaluable stents was 3.03+/-0.31 mm (P=0.0002). Overall, six of 12 in-stent restenoses were correctly detected by MDCT [50% sensitivity (confidence interval 22-77%)] and in 51 of 90 lesions, in-stent restenosis was correctly ruled out [57% specificity (46-67%)]. In evaluable stents, six of seven in-stent restenoses were correctly detected, and the absence of in-stent stenosis was correctly identified in 51 of 52 cases [sensitivity 86% (42-99%) and specificity 98% (88-100%)].

Conclusion: Stent type and diameter influence evaluability concerning in-stent restenosis by MDCT. The rate of assessable stents is low, but in evaluable stents, accuracy for detection of in-stent restenosis can be high.

Citing Articles

Enhancing diagnostic performance and image quality in coronary CT angiography: Impact of SnapShot Freeze 2 algorithm across varied heart rates in stent patients.

Wu Z, Han Q, Liang Y, Zheng Z, Wu M, Ai Z J Appl Clin Med Phys. 2024; 25(8):e14412.

PMID: 38807292 PMC: 11302822. DOI: 10.1002/acm2.14412.


Bioresorbable Magnesium-Based Stent: Real-World Clinical Experience and Feasibility of Follow-Up by Coronary Computed Tomography: A New Window to Look at New Scaffolds.

Ghafari C, Brassart N, Delmotte P, Brunner P, Dghoughi S, Carlier S Biomedicines. 2023; 11(4).

PMID: 37189769 PMC: 10135647. DOI: 10.3390/biomedicines11041150.


Stent imaging on a clinical dual-source photon-counting detector CT system-impact of luminal attenuation and sharp kernels on lumen visibility.

Decker J, ODoherty J, Schoepf U, Todoran T, Aquino G, Brandt V Eur Radiol. 2022; 33(4):2469-2477.

PMID: 36462045 DOI: 10.1007/s00330-022-09283-4.


Stent visualization methods to guide percutaneous coronary interventions and assess long-term patency.

Ghafari C, Carlier S World J Cardiol. 2021; 13(9):416-437.

PMID: 34621487 PMC: 8462039. DOI: 10.4330/wjc.v13.i9.416.


Health Care Monitoring and Treatment for Coronary Artery Diseases: Challenges and Issues.

Alghrairi M, Sulaiman N, Mutashar S Sensors (Basel). 2020; 20(15).

PMID: 32752231 PMC: 7435700. DOI: 10.3390/s20154303.