Clinical Performance of Contemporary Tibial Polyethylene Components
Overview
Affiliations
A postclinical retrieval analysis was performed on 43 polyethylene tibial components of a contemporary total knee arthroplasty system with implantation duration between 12 and 80 months. Components were scored for 8 potential modes of surface wear or damage on the top and back surfaces. Moderate backside wear of 4.1 microm/y was documented by measuring the extent of manufacturer's engraved lettering removal. Neither the topside nor backside score correlated with duration of in vivo function. No component experienced topside or backside delamination, cracking, or significant deformation. The greatest contribution to wear and damage score was pitting and scratching secondary to bone cement debris. The extent of both wear and damage experienced by these components was moderate, in comparison with that previously reported with older implant systems.
A Comparison of Wear Patterns on Retrieved and Simulator-Tested Total Knee Replacements.
Dammer R, Zietz C, Bader R J Funct Biomater. 2022; 13(4).
PMID: 36412896 PMC: 9680524. DOI: 10.3390/jfb13040256.
Ngai V, Kunze J, Cip J, Laurent M, Jacobs J, Wimmer M J Orthop Res. 2020; 38(7):1607-1616.
PMID: 32410286 PMC: 7329356. DOI: 10.1002/jor.24720.
Lapaj L, Mroz A, Kokoszka P, Markuszewski J, Wendland J, Helak-Lapaj C Acta Orthop. 2016; 88(1):62-69.
PMID: 27781667 PMC: 5251266. DOI: 10.1080/17453674.2016.1248202.
Relationship of surface damage appearance and volumetric wear in retrieved TKR polyethylene liners.
Knowlton C, Bhutani P, Wimmer M J Biomed Mater Res B Appl Biomater. 2016; 105(7):2053-2059.
PMID: 27401236 PMC: 5233657. DOI: 10.1002/jbm.b.33684.
Pourzal R, Knowlton C, Hall D, Laurent M, Urban R, Wimmer M Clin Orthop Relat Res. 2016; 474(8):1867-75.
PMID: 26891896 PMC: 4925408. DOI: 10.1007/s11999-016-4749-8.