Objective: To report the initial experiences with the 25-gauge transconjunctival sutureless vitrectomy (TSV) system, including intraoperative and postoperative problems.

Methods: We retrospectively reviewed the medical records and surgical videotapes of 50 consecutive patients who underwent vitrectomy performed by one surgeon using the TSV system.

Results: Intraoperatively, we encountered such problems as difficulty in inserting the microcannula, which led to deformity, instability of the microcannula, self-disconnection of the infusion tip and resultant lens damage, and conversion to 20-gauge conventional vitrectomy. Postoperatively, there were 8 cases with hypotony (IOP < 6 mm Hg) on day 1, 6 cases with elevated IOP, and 3 cases with retinal detachment during follow-up.

Conclusion: Though certain problems exist during and after surgeries using TSV, this system is both convenient and safe for various vitreoretinal procedures.

Citing Articles

A Rare Complication Causing Another Rare Complication: Corneal Dellen with Infiltration Secondary to Subconjunctival Silicon Oil.

Mohan S, Nadri G, Mohan S Beyoglu Eye J. 2023; 8(3):233-236.

PMID: 37766771 PMC: 10521125. DOI: 10.14744/bej.2023.74508.


Intraocular pressure outcomes after 23-G vitreoretinal surgery with two different transconjunctival sutureless sclerotomy techniques: vertical versus tunnel entry.

Seyyar S, Saygili O, Tiskaoglu N Int Ophthalmol. 2022; 42(6):1763-1769.

PMID: 35079939 DOI: 10.1007/s10792-021-02172-7.


Comparison of safety and effectiveness between 23-gauge and 25-gauge vitrectomy surgery in common vitreoretinal diseases.

Sedova A, Steiner I, Matzenberger R, Georgopoulos M, Scholda C, Kriechbaum K PLoS One. 2021; 16(3):e0248164.

PMID: 33667273 PMC: 7935274. DOI: 10.1371/journal.pone.0248164.


Comparative study of straight vs angled incision in 27-gauge vitrectomy for epiretinal membrane.

Yomoda R, Sasaki H, Kogo J, Shiono A, Jujo T, Sekine R Clin Ophthalmol. 2018; 12:2409-2414.

PMID: 30538424 PMC: 6263245. DOI: 10.2147/OPTH.S183456.


Dellen formation as a complication of subconjunctival silicone oil following microincision vitrectomy.

Mahgoub M, Roshdy M, Wahba S Clin Ophthalmol. 2018; 11:2215-2219.

PMID: 29290680 PMC: 5736360. DOI: 10.2147/OPTH.S149531.